Optimization and scale up of production of the PSMA imaging agent [18F]AlF-P16-093 on a custom automated radiosynthesis platform

Abstract

Background

Recent advancements in positron emission tomograph (PET) using prostate specific membrane antigen (PSMA)-targeted radiopharmaceuticals have changed the standard of care for prostate cancer patients by providing more accurate information during staging of primary and recurrent disease. [68Ga]Ga-P16-093 is a new PSMA-PET radiopharmaceutical that demonstrated superior imaging performance in recent head-to-head studies with [68Ga]Ga-PSMA-11. To improve the availability of this new PSMA PET imaging agent, [18F]AlF-P16-093 was developed. The 18F-analog [18F]AlF-P16-093 has been synthesized manually at low activity levels using [18F]AlF2+ and validated in pre-clinical models. This work reports the optimization of the production of > 15 GBq of [18F]AlF-P16-093 using a custom automated synthesis platform.

Results

The sensitivity of the radiochemical yield of [18F]AlF-P16-093 to reaction parameters of time, temperature and reagent amounts was investigated using a custom automated system. The automated system is a low-cost, cassette-based system designed for 1-pot syntheses with flow-controlled solid phase extraction (SPE) workup and is based on the Raspberry Pi Zero 2 microcomputer/Python3 ecosystem. The optimized none-decay-corrected yield was 52 ± 4% (N = 3; 17.5 ± 2.2 GBq) with a molar activity of 109 ± 14 GBq/µmole and a radiochemical purity of 98.6 ± 0.6%. Run time was 30 min. A two-step sequence was used: SPE-purified [18F]F was reacted with 80 nmoles of freeze-dried AlCl3·6H2O at 65 °C for 5 min followed by reaction with 160 nmoles of P16-093 ligand at 40 °C for 4 min in a 1:1 mixture of ethanol:0.5 M pH 4.5 NaOAc buffer. The mixture was purified by SPE (> 97% recovery). The final product formulation (5 mM pH 7 phosphate buffer with saline) exhibited a rate of decline in radiochemical purity of ~ 1.4%/h which was slowed to ~ 0.4%/h when stored at 4 °C.

Conclusion

The optimized method using a custom automated system enabled the efficient (> 50% none-decay-corrected yield) production of [18F]AlF-P16-093 with high radiochemical purity (> 95%). The method and automation system are simple and robust, facilitating further clinical studies with [18F]AlF-P16-093.

Introduction

Positron emission tomography (PET) using prostate-specific membrane antigen (PSMA)-targeted radiopharmaceuticals is transforming clinical care for prostate cancer (PCa) patients by providing physicians with more accurate information during staging of primary and recurrent disease (Fendler et al. 2023; Houshmand et al. 2023). The first PSMA-PET radiopharmaceuticals approved in the United States were Ga68-PSMA-11 and Pylarify®—F18 DCFPyl, followed by kit formulations of PSMA-11 (Illuccix®—Ga 68 gozetotide; Locametz®—Ga 68 gozetotide) (Carlucci et al. 2021; Voter et al. 2022; FDA 2022a, b). These PSMA radiotracers have shown similar clinical utility in both head-to-head studies (Dietlein et al. 2017; Dietlein et al. 2015) and retrospective analyses (Evangelista et al. 2022), including a recent pilot study that concluded that both [68Ga]Ga-PSMA-11 and [18F]FPyl may be suitable for selecting patients for radioligand therapy (RLT) using Pluvitco® (lutetium Lu 177 vipivotide tetraxetan, also known as [177Lu]Lu-PSMA-617; Fallah et al. 2023) using the same thresholding criteria (Heilinger et al. 2023).

In these studies, however, differences in image contrast between 68Ga- and 18F-PSMA-PET tracers were reported. For example, Heilinger et al. (2023) observed a twofold increase in the median contrast to noise ratio of [18F]DCFPyl over [68Ga]Ga-PSMA-11 (Heilinger et al. 2023). The authors surmised that improved image contrast was due in part to the later imaging (120 min after injection) which was possible because of the longer half-life of fluorine-18 compared to gallium-68 (109.6 min vs. 68 min) and higher injected activity (Dietlein et al. 2015; Heilinger et al. 2023). In general, higher resolution images are obtained with fluorine-18 vs gallium-68 due to the differences in their physical decay properties, the former having a higher branching fraction and lower average positron energy (Carter et al. 2020).

Given these advantages of 18F-PSMA, combined with the existence of a cyclotron-based supply chain providing high-activity batches of fluoride-18 throughout much of the developed world, fluoride-18 would be the preferred radionuclide for the development of any new PSMA-PET imaging agents. Radiotracers based on fluoride-18 can be produced in a centralized pharmacy and distributed widely, each batch supporting many more patients than possible with gallium-68, which is often limited to 2–4 patients/generator. Though cyclotron-produced gallium-68 is also available (Thisgaard et al. 2021), it is not yet widely available and the shorter half-life of gallium-68 limits distribution of the final product. Hence, the latest PSMA-PET agent to receive regulatory approval by the FDA in 2023, Posluma® (rh PSMA 7.3), utilizes fluoride-18 as the radionuclide for the imaging molecule in a radiohybrid design (Jani et al. 2023; Surasi et al. 2023). PSMA-1007, also an 18F-based tracer is in widespread clinical development though not commercially available. Its pharmacokinetics was designed to minimize background/spillover from bladder activity, a potential advantage in detection of pelvic lesions in recurrent PCa disease (Wurzer et al. 2020; Eiber et al. 2020).

[68 Ga]Ga-P16-093 is a second-generation PSMA-PET agent in clinical trial (Lee et al. 2022; Green et al. 2020; Bahler 2023). It has demonstrated an altered excretion pathway in humans leading to significantly lower bladder activity (Green et al. 2020; Wang 2023) as well as improved tumor detection characteristics compared to [68Ga]Ga-PSMA-11 in the same patient, particularly in individuals with diagnosis of a low/intermediate risk PCa (Wang 2023). Given the aforementioned advantages of fluoride-18, a 18F-based tracer would be preferred, and [18F]AlF-P16-093 was described recently (Zha et al. 2021). [18F]AlF-P16-093 was synthesized manually in high yield using an adaptation of the method first described by McBride et al. (2009) based on the chemical species [18F]AlF2+. [18F]AlF-P16-093 exhibited excellent specific binding profiles in cell, tissue and mouse tumor systems and was considered a good candidate for evaluation in humans (Zha et al. 2021).

Before a novel PET-agent can be investigated routinely in a clinical setting, large-scale synthesis must be demonstrated, preferably using an automated platform. [18F]AlF-P16-093 incorporates the acyclic chelator HBED-CC (see Fig. 1), like PSMA-11, which forms the basis for application of the [18F]AlF2+ method. Several groups have demonstrated successful large-scale, automated production of [18F]AlF-PSMA-11 using [18F]AlF2+ in moderate yield (< 25% not decay corrected), high radiochemical purity, and good stability in high activity batches (> 10 GBq) (Kersemans et al. 2018; Giglio et al. 2018). The goal of this work was to scale up and automate the production of [18F]AlF-P16-093, guided in part by these previous successes with [18F]AlF-PSMA-11. We report here a procedure for the large-scale production of [18F]AlF-P16-093 in high yield (> 50% not decay corrected) using a custom automated synthesis platform.

Fig. 1
figure 1

Reaction scheme with chemical structures of P16-093 and [18F]AlF-P16-093

Methods

General

Several important chemicals were obtained commercially; AlCl3 hexahydrate (ReagentPlus®), sodium acetate trihydrate (BioUltra grade) and water (ACS reagent, for ultratrace analysis) were purchased from Sigma-Aldrich (Milwaukee, WI). Acetic acid (Certified ACS) was bought from Thermo Fisher Scientific (Fair Lawn, NJ). TLC Silica gel 60 F254 plates were obtained from Merck KGaA, Germany. Sep-Pak© Plus Light QMA (Cl-), Sep-Pak© Plus Light QMA and Oasis® HLB 6 cc/150 mg cartridges were purchased from Waters (Milford, MA). High-performance liquid chromatography (HPLC) analysis was performed on an Agilent 1100 series (Agilent, USA) equipped with NaI (TI) scintillation detector (Model 106, Ecker-Ziegler, Germany) with a reversed-phase column (Luna, C18, 150 × 4.6 mm, 5 micron, Phenomenex, USA). Lyophilization was carried out using a SP VirTis Genesis Pilot Freeze Dryer (Warminster, PA). F18 Activity was measured using a Packard Cobra II auto gamma counter (Perkin Elmer, Waltham, MA) or a Capintec (Floral Park, NJ) Model CRC 127-R dose calibrator. Free ligand, P16-093, was synthesized as described previously (Zha et al. 2018).

[18F]fluoride was produced with the IBA cyclotron, Cyclone 18/9 (Louvain-La-Neuve, Belgium), via the 18O (p, n) 18F reaction at the University of Pennsylvania with a niobium target and PEEK delivery lines. For “18F-rinse” studies, [18F]fluoride was obtained by rinsing the target and delivery lines with 1 mL of [18O]O-water after a previous production run. [18O]O-water (> 97%) was purchased from Huayi Isotopes (Changshu, China) or ABX (Radeberg, Germany).

Stopcock manifolds (polypropylene, PT-188, Huayi Isotopes) for automated synthesis were purchased from Nucmedcor (San Francisco, CA). Flexelene® and silicone (Pharma 50) tubing for synthesis cassettes were purchased from Eldon James (Fort Collins, CO) and Cole-Parmer (Vernon Hills, IL) respectively. The reaction vessel for automated synthesis was made of a Wheaton® (Millville, NJ) 10 mL E-Z EX-TRACTION™ vial, a threaded (22–400) phenolic cap (DWK, Millville, NJ) and a silicone septum with PTFE lining (20 × 3 mm) purchased from Tisch Scientific (Cleveland, OH).

Statistical data comparison was done using a Student’s T-Test (Microsoft Excel).

Custom automated radiosynthesis platform

The automated radiosynthesis system (PiBOX) is a cassette-based, 1-pot device with solid phase extraction (SPE) workup capability (see Fig. 2). A typical configuration includes 20 actuators controlling 4 single-use, commercially available cassettes, each made of 5 ganged stopcocks. It was designed specifically for 1-pot reactions that do not require HPLC purification (e.g. Yao et al. 2019; Zhao et al. 2019), and supports standard 18F-recovery and drying methods typical of 18F-displacement reactions. A design goal was more consistent removal of chemical impurities using flow controlled SPE workup. PiBOX includes 4 linear actuators for flow-controlled rinsing and recovery of the product from an SPE cartridge using standard sterile, disposable 5–20 mL syringes. PiBOX is based on the Raspberry Pi Zero 2 W microcomputer (https://www.raspberrypi.com) and uses three commercially available Raspberry Pi HATs (hardware attached on top) to interface the servomotors for stopcock control and the reactor’s heater and thermocouple. Configuration, control and sequence generation software was programmed in the Python3 programming language (https://www.python.org/download/releases/3.0/). Figure 2 shows the fluidic hardware configuration of PiBOX for this work. Each synthesis utilizes a disposable set of 4 stopcock manifolds with associated tubing and fittings, glass vessels, sterile syringes and SPE cartridges.

Fig. 2
figure 2

PiBOX cassette setup for production of [18F]AlF-P16-093. 1. 10 mL ultrapure water; 2. 6 mL water; 3. 10 mL water; 4. 10 mL saline; 5. 3 mL 2:1 ethanol: phosphate buffer; 6. 0.7 mL 4:1 pH 4.5 0.5 M NaOAc buffer: ethanol; 7. P16-093 in 0.45 mL ethanol; 8. formulation vial with 15.8 mL phophate buffer and 1.2 mL 14% saline. The black rectangles above syringes 3–5 represent linear actuators used for flow controlled SPE rinsing and elution

PiBOX yield optimization using 18F-rinses

Experiments were carried out using three different newly constructed PiBOX machines to determine the effects of important reaction parameters including reagent amounts, time, temperature and reagent sequencing on the overall radiochemical yield in the following manner. 2 mL of ultrapure water was introduced into a 5 mL conical glass vial that was rinsed previously with 1–2 mL ultrapure water before every experiment. The entire 18F-rinse volume (~ 0.5–1 mL; ~ 0.2–2 GBq) was transferred by hand using a metal needle and syringe and added to the conical vial. 18F-activity in the vial was assayed using an ion chamber before execution of the automated radiosynthesis. The following steps were then carried out automatically. The 18F-activity was transferred by vacuum through a Sep-Pak© Light QMA or QMA carb cartridge that had been conditioned with 10 mL of pH 4.5 0.5 M NaOAc buffer followed by 10 mL ultrapure water. The cartridge was rinsed with 10 mL of trace metal grade water then purged with air for 30 s by vacuum. The QMA cartridge was rinsed with 0.7 mL of 20–33% ethanol in pH 4.5 0.5 M NaOAc buffer to elute the trapped 18F activity into the reaction vessel by applying vacuum. Previously, the reaction vessel was loaded with AlCl3·6H2O (20–160 nmoles; 2–16 µL; 10 mM AlCl3·6H2O). In some experiments the AlCl3·6H2O was freeze-dried inside the bottom of the reaction vessel by aliquoting 2–16 µL AlCl3·6H2O (10 mM) into a reaction vessel and placing it into the freeze drier. After freeze drying, the AlCl3·6H2O-containing reaction vessels were stored at room temperature until use. After elution of the 18F-activity from the QMA into reaction vessel containing either the freeze-dried or liquid AlCl3·6H2O, the solution was heated for 5 min @ 50–65 °C then cooled for 30 s using compressed gas to ~ 35–45 °C. P16-093 ligand (80–160 nmoles) in 0.45 mL ethanol was added to the cooled solution and heated for 2–5 min @ 35–65 °C. In some experiments P16-093 was dissolved in 0.35 mL ethanol and 0.1 mL pH 4.5 0.5 M NaOAc buffer. In all cases the final P16-093 solution was ~ 1:1 pH 4.5 0.5 M NaOAc buffer:ethanol when combined with the QMA eluent. After heating the reaction mixture was cooled for 60 s to ~ 25–30 °C. This sequence of reagent addition and heating was named “two step” as [18F]AlF2+ is presumably formed first (Step 1) before reaction with P16-093 (Step 2).

For a subset of experiments radiolabeling was carried out using a “one-step” method where the 18F-activity eluted from the QMA and the AlCl3·6H2O were not heated separately from the P16-093. Instead, P16-093 in 0.45 mL ethanol was added to the reaction vessel immediately after elution of the QMA so that the [18F]F, AlCl3·6H2O and P16-093 were all heated together for 5–10 min @ 40–50 °C. After heating the mixture was cooled for 60 s using compressed gas to ~ 25–30 °C.

After labeling (either 1 or 2 step) the reaction mixture was purified by SPE using a single HLB 6 cc 150 mg cartridge. The HLB cartridge was conditioned before use with 10 mL ethanol followed by 10 mL water. The reaction mixture was first diluted with 6 mL of water then transferred using pressure onto and through the HLB cartridge. The cartridge was rinsed in the manner suggested by Kersemans et al. (2018), first with 10 mL saline then 10 mL water using 2 separate linear actuator driven syringes @ ~ 10 mL/min. The product was eluted form the HLB with 3 mL of 2:1 ethanol 5 mM phosphate buffer (pH 7–7.5) into a vial containing 15.8 mL 5 mM phosphate buffer (pH ~ 7) with 1.2 mL concentrated (14%) saline to adjust osmolality. The final solution, 20 mL, is sterilized by passing it through a Millex GV filter using pressure, though for optimization and scale up reactions in this work no filter was used and the solution was transferred directly into a product vial.

At the end of an automated synthesis, the following components were assayed for 18F-activity using a dose calibrator: product vial (P), HLB cartridge (HLB), HLB waste (combined rinse and diluted reaction mixture volumes—HLBw), formulation vial (FV), QMA cartridge (QMA), QMA waste (combined rinse and 18F-conical vial volumes—QMAw), reaction vessel (RxV) and the 18F-conical vial (CVr).

For comparisons between runs a yield (RCY) was calculated using Eq. 1 below. All measurements were decay corrected to the time of measurement of the initial 18F-activity in the conical vial (CVi) at the start of synthesis.

$$RCY\left(\%\right)=\frac{P}{{CV}_{i}-{CV}_{r}}\times 100$$
(1)

A yield based on the activity eluted from the QMA was calculated using Eq. 2 to remove any effects of changing efficiency of 18F-recovery from QMA on the RCY. All activities were decay corrected to the time of measurement of starting [18F]fluoride in the conical vial.

$${RCY}_{QMA}=\frac{P}{{CV}_{i}-{CV}_{r}-QMA-{QMA}_{w}} \times 100$$
(2)

Finally, all the decay corrected component activities were summed, normalized to the starting 18F-activity and then expressed as a percent to determine the 18F-activity recovery (AR).

For large scale production ([18F]F > 30 GBq) the initial activity (CVi) was not measured so the RCY was calculated using the average AR from the 18F-rinse runs and the following formulas:

$$ALL=P+HLB+{HLB}_{w}+FV+QMA+{QMA}_{w}+{R}_{x}V+{CV}_{r}$$
(3)
$$RCY.LS (\%)=\frac{P}{(\frac{ALL}{AR}-{CV}_{r})} \times 100$$
(4)
$$RCY.{LS}_{QMA}(\%)=\frac{P}{(\frac{ALL}{AR}-{CV}_{r}-QMA-{QMA}_{w})}\times 100$$
(5)

The efficiency of QMA extraction was calculated by using the formula below for all combinations of QMA elution solution and QMA type:

$${QMA}_{eff}(\%)=\frac{{CV}_{i}-{CV}_{r}-QMA-{QMA}_{w}}{{CV}_{i}-{CV}_{r}}\times 100$$
(6)

Time dependence of 18F-incorporation

To investigate directly if there was a time dependence (2–10 min) of 18F-incorporation into P16-093 under representative conditions of the PiBOX “two step” yield optimization studies the following experiment was carried out. Three PiBOX reaction vessels were prepared each containing 0.5 mL of 20% ethanol: pH 4.5 0.5 M NaOAc and 100 μg (80 nmoles) of P16-093 in 0.45 mL ethanol. [18F]AlF2+ was prepared automatically with PiBOX as described above using a reaction vessel with 120 nmoles of AlCl3·6H2O. A QMA carb cartridge was used to purify 18F-activity and 0.7 mL of 20% ethanol: pH 4.5 0.5 M NaOAc was used to elute the QMA cartridge. The mixture of buffer and AlCl3·6H2O was heated for 5 min @ 65 °C then cooled for 30 s to ~ 40 °C. The PiBOX sequence ended, and the reaction vial containing ~ 2 GBq [18F]fluoride was manually split into three equal aliquots (~ 200 μL). Each aliquot was added to one of the 3 previously prepared reaction vessels which were then placed in a preheated oil bath (40 °C) and sampled at 2, 5 and 10 min. Previously it was verified that the internal liquid temperature during PiBOX heating reached ~ 40 °C in 2–3 min. The samples were diluted with 10% ethanol with phosphate buffer (5 mM pH ~ 7) and put on ice until analyzed by both TLC and HPLC. The percent of total 18F-activity in the sample present as [18F]AlF-P16-093 was determined.

Large scale production of [18F]AlF-P16-093

Radiosynthesis of [18F]AlF-P16-093 was carried out on a PiBOX machine (PiBOX-1) installed in a hot cell at the cyclotron facility at the University of Pennsylvania using the same program sequence and the same lots for reagents as in the companion low activity 18F-rinse runs. Three runs where carried out for each of two different “optimized” (reagents, time, temperature) conditions and the RCY’s and RCYqma’s from the low (18F-rinse) and high activity (~ 37 GBq [18F]fluoride @ cyclotron) production were compared to see if there was an effect of scaling up activity on radiochemical yield using a Student’s T-test. Tests using more than 37 GBq [18F]fluoride were not carried out. 37 GBq was considered an appropriate level to support on-site clinical investigations while minimizing exposure for this work, as all synthesis components were assayed for 18F-activity the same day for radiochemical yield and other efficiency estimates.

Stability of final product

The stability of the final product solution was determined by sampling the product vial for up to 5 h after end of synthesis and analyzing the 18F-activity by thin layer chromatography (TLC) or HPLC. Stability was determined for storage at room temperature (RT) and 4 °C for representative low-level 18F-syntheses starting with 160 nmoles P16-093, and for 3 large scale production runs using 120 nmoles (RT only).

Storage of reagents

P16-093 starting material was stored as either individual aliquots (100–200 µg; 80–160 nmoles) for each run at − 20 °C, or as an ethanol solution (2 mg/mL) stored at − 20 °C. Aliquots were prepared by making a 1 mg/mL solution of P16-093 in ethanol then removing the solvent from 100 to 200 µL aliquots. AlCl3·6H2O (10 mM) stock solution was stored at 4 °C. For experiments using freeze-dried AlCl3·6H2O, aliquots (2–16 µL) were added directly to the PiBOX reaction vessel and the solvent removed using a freeze drier. The freeze-dried vessels were stored at RT.

Results

PiBOX yield optimization studies using 18F-rinses

Based on published conditions for large scale production of [18F]AlF-PSMA-11 (Kersemans et al. 2018) and experience with the manual preparation of [18F]AlF-P16-093 (Zha et al. 2021), initial conditions for labeling were set at 80 nmoles AlCl3·6H2O with ~ 160 nmoles P16-093 (200 µg) using 65 °C for 5 min for both [18F]AlF2+ formation and P16-093 labeling carried out in two separate steps. Table 1 shows the results from a series of experiments carried out from October 2022 through July 2023 using 3 different PiBOX machines under varying reaction conditions and reagent preparations. Table 1 suggests a trend towards higher yields using freeze dried AlCl3·6H2O with more mild (lower temperature and shorter time) P16-093 labeling conditions. For example, the average RCYqma using liquid AlCl3 for runs 1–8 excluding run 6 (56 ± 1.9%) was statistically different (p = 0.00004) than the average RCYqma of freeze-dried AlCl3 runs 9–11 (66 ± 1.6%). The maximum yield was obtained using 30 – 40 °C for labeling P16-093 for 2–5 min using the two-step method. A similar high yield (79 ± 1.5%) was observed using a one-step method (40–50 °C for 10 min; runs 21, 23 and 25), which was not statistically different (p = 0.54) when compared to the average yield for runs 17–19 (78 ± 2.1%).

Table 1 Results from “18F-rinse” optimization studies using 80 nmoles AlCl3·6H2O with ~ 160 nmoles P16-093

Based on the preliminary optimization results in Table 1, a standard method for all PiBOX devices was chosen for further optimization studies. The standardized procedure consisted of a two-step synthesis using 65 °C for 5 min for [18F]AlF2+ formation followed by labeling of P16-093 for 4 min at 40 °C. Freeze-dried AlCl3·6H2O and P16-093 aliquots (100–200 µg) that were stored as a solid at -20 °C were used for the standardized method, and 1:4 ethanol/0.5 M NaOAc (pH 4.5) was used to elute activity from the QMA cartridge.

Figure 3 summarizes the results from experiments designed to determine the sensitivity of PiBOX RCYqma to changes in starting P16-093 amount and relative amount of AlCl3·6H2O using the standardized conditions. There was a small but statistically significant decrease in yield for the 2:1 mol ratio of P16-093: AlCl3·6H2O when comparing 200 µg to 100 µg (P < 0.05), and also a trend towards lower yields when the mole ratio of P16-093: AlCl3·6H2O was decreased from 2:1 to ~ 1:1 using the same P16-093 mass. Using 200 µg P16-093, the decrease in yield from 2:1 to 1:1 was statistically significant (P < 0.02). Finally, at higher mole ratios (4:1), data from Fig. 3 suggests a trend (P = 0.06) towards slightly lower yields using 200 µg (160 nmol), but a significant (P < 0.0001) drop off for 100 µg (80 nmol) where the yield using a ratio 4:1 was only 13.7 ± 1.1% compared to 61 ± 4.6% using a mole ratio of 2:1. This large decrease in yield was not observed when only ~ 1/10 of the starting 18F-rinse activity (~ 55 MBq) was used with the exact same conditions and reagents (80 nmoles P16-093; 20 nmoles AlCl3·6H2O) where the yield was 62.4% in a single experiment (bar graph in Fig. 3).

Fig. 3
figure 3

Effects of starting P16-093 and mole ratio of P16-093: AlCl3·6H2O on yield using optimized PiBOX reaction conditions. N = 3–4 for each data point. *P < 0.05; **P < 0.02; ***P < 0.0001. The bar graph represents results from a single experiment using the same conditions and reagents as the 4:1 mol ratio (80 nmol P16-093) runs except that the 18F-rinse volume was diluted with ultrapure water and sampled so that ~ 1/10 of the original 18F-rinse activity and volume was used for PiBOX

Time dependence of P16-093 labeling after [18F]AlF2+ formation

The effect incubation time of [18F]AlF2+ with P16-093 on 18F-incorporation @ 40 °C is illustrated in Fig. 4. There was no difference in the percent of 18F-activity that was present as [18F]AlF-P16-093 from 2 to 10 min for two independent reactions measured by either HPLC or TLC. Only 2 of the 3 separate reaction vials prepared were used due to technical problems (1 vial lost to oil bath).

Fig. 4
figure 4

Time dependence of 18F-incorporation after incubation (40 °C) of 100 µg P16-093 (0.45 mL ethanol + 0.5 mL 20% ethanol in pH 4.5 0.5 M NaOAc buffer) with an aliquot (~ 1/3) of a reaction vessel that was reacted with QMA-purified [18F]fluoride and 120 nmoles AlCl3·6H2O using PiBOX. The final ratio of P16-093:AlCl3·6H2O was ~ 2:1 in each of 3 vials used for the time dependence study

Large scale production of [18F]AlF-P16-093

Table 2 shows the results of the large scale (> 30 GBq [18F]F) production runs using the optimized PiBOX conditions. There was no difference in the average RCYqma of the large-scale runs compared with the companion (same lot for all reagents) low-level 18F-rinse runs. For example, for the mole ratio of 2:1 P16-093: AlCl3·6H2O with 200 µg P16-093, the 18F-rinse runs gave a yield of 72 ± 4.7% while the scaled up runs using the same reagents had an average yield of 72 ± 4.3% (P = 0.97). A similar result was obtained when comparing low-level fluoride-18 runs using a mole ratio of 1.5:1 (150 µg P16-093) with the associated high 18F-activity runs: 66 ± 7.3% versus 66 ± 6.0% respectively. Finally, there were no differences in starting activity or yield between the two sets of high activity runs (Table 2), but there was a significant (P < 0.05) increase in molar activity from 110 ± 14 GBq/µmole to 140 ± 11 GBq/µmole when less P16-093 was used.

Table 2 Results from scaled up production runs using optimized conditions

Stability of [18F]AlF-P16-093 preparations

Radiochemical purity determined by TLC as a function of time is ploted in Fig. 5 for room temperature (RT) and refrigerator (~ 4 °C) storage conditions (left panel). The data was compiled from 5 different low level 18F-optimization runs on PiBOX using 200 μg of P16-093 and 80 nmoles AlCl3·6H2O. In two runs the product vial was split at the end of synthesis and stored at two different temperatures. In other experiments 100% of the product vial was stored at RT (N = 2) or in a refrigerator (N = 1). Data from each sample time point for all experiments was averaged and plotted together. Rates of loss of [18F]AlF-P16-093 were determined to be ~ 1.4%/h and ~ 0.4%/h for RT and refrigerator storage conditions respectively using the slopes of the linear trend lines shown in Fig. 5 left panel. The radiochemical purity as a function of time stored at RT is shown in Fig. 5 right panel for the 3 high activity runs 4–6 listed in Table 2. The average decrease in radiochemical purity was 1.34 ± 0.07%/h (average of the 3 individual trend line slopes).

Fig. 5
figure 5

Left plot: summary of stability data of [18F]AlF-P16-093 stored at RT (square) and ~ 5 °C (circle) from low F18 activity optimization runs (N = 3–4) using 200 μg P16-093 and 80 nmoles AlCl3·6H2O. Right plot: individual run stability data from RT storage of [18F]AlF-P16-093 from 3 high activity (product > 15 GBq) production runs using 150 μg P16-093 and 80 nmoles AlCl3·6H2O. Product formulation is 10% ethanol in 5 mM phosphate buffer ~ pH 7

QMA and HLB extraction efficiencies

During 66 optimization runs 2 different QMA types were used with 2 different elution solutions. Table 3 summarizes the efficiencies for each case.

Table 3 Results from optimization runs using 3 different PiBOX machines

After QMA purification of [18F]F by QMA SPE, reaction of [18F]F with AlCl3·6H2O and 18F-labeling of P16-093, purification was carried out by flow-controlled SPE using an HLB 6 cc (150 mg) cartridge similar to the method described by Kersemans et al. (2018). The average 18F-activity (% of total starting activity) remaining on HLB was 1.7 ± 0.5% (N = 67) for the optimization runs. The waste from the HLB was assayed for [18F]AlF-P16-093 by TLC (0.6 ± 0.2%; N = 6). For the six runs where the HLB waste was assayed for [18F]AlF-P16-093, an HLB efficiency for recovery of [18F]AlF-P16-093 calculated to be 97.5 ± 1.1%, assuming all the activity remaining on the HLB was [18F]AlF-P16-093.

Discussion

[18F]AlF-P16-093 seeks to leverage the simple and high yield method of [18F]AlF2+ (McBride et al. 2009) to convert [68Ga]Ga-P16-093 into an 18F-analog that is suitable for wider clinical use while maintaining the imaging advantages demonstrated by [68Ga]Ga-P16-093 (Wang et al. 2023). Though previously synthesized manually for pre-clinical evaluation (Zha et al. 2021), [[18F]AlF-P16-093 has not been systematically optimized on an automated platform and scaled up for “clinical-level” production (e.g., 37 GBq 18F]F).

Scaling up and/or automating radiochemical reactions based on manual syntheses is not always straight forward (Kersemans et al. 2018; Wurzer, et al. 2021). In particular, manual reactions using [18F]AlF2+ for chelation labeling (AlF) or [18F]F for isotopic exchange (IE) reactions typically fractionate 18F-activity from the target such that only a small proportion of the target volume (and associated impurities like metal ions) is used. Combined with very small reagent volumes (2–20 uL), final reaction volumes in manual [18F]AlF2+ reactions can be very small (0.2–0.5 mL). These attributes are often not amendable to traditional automation and lower yields are not uncommon when scaling up with machines (Kersemans et al. 2018; Wurzer et al. 2021).

Manual reactions that fractionate [18F]F may also not represent the true levels of [19F]fluoride present. While 37 GBq of [18F]fluoride is only ~ 0.5 nmoles based on the radionuclide’s theoretic specific activity, practical molar activities of final products suggest that [19F]fluoride mass is 5–50 nmoles or more (Cleeren et al. 2016; Link et al. 2012), as [19F]fluoride is known to contaminate [18F]fluoride production from several sources including delivery lines and reagents (Link et al. 2012). This amount of [19F]fluoride can be significant proportion of reagent amounts in AlF and IE reactions where precursor amounts are often 100–200 nmoles. Cleeren et al. (2016) demonstrated this directly by titrating AlF reactions with [19F]fluoride under standard AlF labeling conditions (40 nmoles AlCl3; 150 nmoles ligand), showing a decrease in RCY when > 40 nmoles [19F]F is added. Wurzer et al. (2021) calculated that [19F]fluoride can be present at > 10 nmoles just from typical target water volumes based on known specifications, and therefore [18F]F molar activity may impact IE RCY directly where only 150 nmol precursor is used (Wurzer et al. 2021).

Accordingly, AlF optimization studies should be carried-out using the same production conditions as in clinical production  to reproduce impurity and [19F]fluoride concentrations that can affect RCY. This is not always practical and, in this work, 100% of “18F-rinse” volumes were used instead for optimization. 18F-rinses (H218O) used about ½ the normal target volume and traversed the same radionuclide delivery path including delivery lines, target body and other standard plumbing shortly after a clinical production run. The RCY of large-scale runs were not different than 18F-rinse optimization runs using 80 nmoles of AlCl3. At lower AlCl3 levels (20 nmoles with 80 nmol P16-093), we observed a significant decrease in RCY that was not observed when an 18F-rinse was “fractionated” (Fig. 3 bar graph). Given that the RCY was high using 80 nmol P16-093 with 40 nmol AlCl3, we attributed the severe decrease in yield to [19F]F- and/or impurities such as metal ions present in the 18F-rinse that competed or otherwise interfered with AlCl3 when only 20 nmoles was used. We concluded that the minimum amount of AlCl3 for high yield under our large-scale production conditions was ~ 40–80 nmoles. Only 80 nmoles was used to validate large-scale production, though Fig. 3 results suggest that 40 nmoles may also work with only a small decline in RCY. Moreover, the effect of aliquoting [18F]fluoride (Fig. 3 bar graph) validated our approach to use 100% of an [18F]F-rinse as a better proxy for real production impurity/[19F]F-concentrations, though greatly limiting throughput. (1 rinse-run/day).

A precursor-to-AlCl3 mole ratio range of ~ 2:1 gave the optimal RCY and is similar to literature results for [18F]AlF-PSMA-11 (Kersemans et al. 2018). Slightly lower mole ratios also gave high yields and molar activity was improved significantly using 120 nmoles of P16-093 with 80 nmoles of AlCl3 (see Table 4). No difference in RCY or radiochemical purity was observed using a one-step versus a two-step reaction during optimization. Though some investigators report that a two-step reaction is necessary for high yields and/or high purity using [18F]AlF2+ with PSMA-11 (Kersemans et al. 2018), this result is consistent with other [18F]AlF-PSMA-11 reactions (Malik et al. 2015; Al-Momani et al. 2017).

Table 4 Large-scale [18F]AlF2+ labeling of acyclic chelators conjugated with an urea-based PSMA inhibitor

QMA (chloride form) was the preferred type (Table 3) compared to QMA (carb), though the difference was very small, particularly when the 20% ethanol eluent was used. In general, recovery of activity from the QMA using 0.7 mL of 20% ethanol in 0.5 M NaOAc in the optimization runs was consistent with published values that use 0.5–0.6 mL of 100% 0.5 M NaOAc (Kersemans et al. 2018; Giglio et al. 2018).

Some investigators report that purification of [18F]AlF-PSMA-11 by SPE was not possible due to excessive breakthrough during automated synthesis and HPLC was used instead (Giglio et al. 2018). Kersemans et al. (2018) also described significant (> 25%) breakthrough using many standard types of SPE bonded phases and proposed using 2 SPE cartridges (HLB 360 mg) in tandem to get > 98% retention. In this work, a single, larger format HLB (6 cc; 150 mg) was sufficient for nearly quantitative recovery using flow-controlled rinsing. Ethanol content of the reaction mixture after dilution before loading the SPE cartridge was ~ 8% and was similar to Kersemans et al. (2018).

The radiochemical purity of the final product remained > 90% when stored at RT for ~ 4 h and exhibited a rate of decline in radiochemical purity similar to Kersemans et al. (2018) for [18F]AlF-PSMA-11 using a similar formulation. Refrigeration slowed decomposition (loss of 18F-label) by about threefold to ~ 0.4%/h. The decline in radiochemical purity over several hours could limit patient throughput. This can be compensated in part by refrigeration, where 14 patients could be done a single scanner while maintaining purity >  = 95% (assuming starting purity is 98%) where patients are scanned every 30 min after 60 min is allowed for QC, packaging and delivery—all refrigerated. This number would reduce to 9 patients/scanner if the dose was at room temperature during QC, packaging and delivery, but otherwise kept refrigerated, and only 3 patients/scanner if the dose was never refrigerated. Therefore, [18F]AlF-P16-093 (like [18F]AlF-PSMA-11) is most suited for local cyclotron-based hospital use and would require refrigeration for limited distribution.

The results of the large-scale [18F]AlF-P16-093 syntheses using PiBOX are compared with high activity production of similar compounds (acyclic chelators conjugated with a PSMA inhibitor) in Table 4. The yield not decay corrected was ~ 2 × higher than for [18F]AlF-PSMA-11 (Kersemans et al. 2018; Giglio et al. 2018) and Glu-NH-CO-NH-Lys(Ahx)L3 (Cleeren et al. 2016) with similar execution time using similar or less starting material. Given that many conditions used here were based in part on published methods for [18F]AlF-PSMA-11, this result was unexpected. The only known differences in method with automated [18F]AlF-PSMA-11 was the use of ethanol/NaOAc mixture to elute the QMA and the use of freeze-dried AlCl3·6H2O. Although we observed a benefit of freeze-dried AlCl3·6H2O, it was not large enough to explain the twofold higher yield. In general, reaction times and temperatures were similar to those used for labeling with commercial machines, although [18F]AlF-PSMA-11 methods that used a 2-step method used RT for reaction of [18F]F with AlCl3 (Kersemans et al. 2018; Giglio et al. 2018) compared to 65 °C used here. Given that conversion of [18F]F to [18F]AlF2+ was reported to be > 80% by Kersemans et al. (2018), this also likely does not explain the difference in overall yield.

Conclusion

The optimized production of [18F]AlF-P16-093 facilitates further clinical implementation and wider adaptation. Its simple, efficient (> 50% none-decay-corrected) labeling at > 15 GBq levels, and desirable performance characteristics in PCa patients makes [18F]AlF-P16-093 an attractive alternative to existing PSMA-targeted tracers. The PiBOX system, made mostly from off-the-shelf, commercially available hardware and opensource software, is a robust and affordable platform for automated [18F]AlF-P16-093 production, may also be suitable for other automated radiochemical syntheses requiring only SPE purification.

Availability of data and materials

Most data generated or analysed during this study are included in this published article. The complete datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AlF:

[18F]AlF2+ for chelation labeling

HPLC:

High performance liquid chromatography

IE:

[18F]F for isotopic exchange

PET:

Positron emission tomography

PCa:

Prostate cancer

PSMA:

Prostate specific membrane antigen

RT:

Room temperature

SPE:

Solid phase extraction

TLC:

Thin layer chromatography

References

  • Al-Momani E, Israel I, Samnick S. Validation of a [Al(18)F]PSMA-11 preparation for clinical applications. Appl Radiat Isot. 2017;130:102–8.

    Article  CAS  PubMed  Google Scholar 

  • Bahler CD, et al. Assessing extra-prostatic extension for surgical guidance in prostate cancer: comparing two PSMA-PET tracers with the standard-of-care. In: Urologic oncology: seminars and original investigations. 2023. Elsevier.

  • Carlucci G, et al. (68)Ga-PSMA-11 NDA approval: a novel and successful academic partnership. J Nucl Med. 2021;62(2):149–55.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • Carter LM, et al. The impact of positron range on PET resolution, evaluated with phantoms and PHITS Monte Carlo simulations for conventional and non-conventional radionuclides. Mol Imaging Biol. 2020;22(1):73–84.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • Cleeren F, et al. New chelators for low temperature Al18F-labeling of biomolecules. Bioconjug Chem. 2016;27(3):790–8.

    Article  CAS  PubMed  Google Scholar 

  • Dietlein M, et al. Comparison of [(18)F]DCFPyL and [(68)Ga]Ga-PSMA-HBED-CC for PSMA-PET imaging in patients with relapsed prostate cancer. Mol Imaging Biol. 2015;17(4):575–84.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • Dietlein F, et al. PSA-stratified performance of (18)F- and (68)Ga-PSMA PET in patients with biochemical recurrence of prostate cancer. J Nucl Med. 2017;58(6):947–52.

    Article  CAS  PubMed  Google Scholar 

  • Eiber M, et al. (18)F-rhPSMA-7 PET for the detection of biochemical recurrence of prostate cancer after radical prostatectomy. J Nucl Med. 2020;61(5):696–701.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • Evangelista L, et al. [68Ga]Ga-PSMA versus [18F]PSMA positron emission tomography/computed tomography in the staging of primary and recurrent prostate cancer: a systematic review of the literature. Eur Urol Oncol. 2022;5(3):273–82.

    Article  PubMed  Google Scholar 

  • Fallah J, et al. FDA approval summary: lutetium Lu 177 vipivotide tetraxetan for patients with metastatic castration-resistant prostate cancer. Clin Cancer Res. 2023;29(9):1651–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • FDA Approves New 68Ga Kit for Prostate Cancer PET. J Nucl Med. 2022a;63(2):26N.

  • FDA Approves Pluvicto/Locametz for Metastatic Castration-Resistant Prostate Cancer. J Nucl Med. 2022b;63(5):13N.

  • Fendler WP, et al. PSMA PET/CT: joint EANM procedure guideline/SNMMI procedure standard for prostate cancer imaging 20. Eur J Nucl Med Mol Imaging. 2023;50(5):1466–86.

    Article  PubMed  PubMed Central  Google Scholar 

  • Giglio J, et al. Synthesis of an Al(18)F radiofluorinated GLU-UREA-LYS(AHX)-HBED-CC PSMA ligand in an automated synthesis platform. EJNMMI Radiopharm Chem. 2018;3(1):4.

    Article  PubMed  PubMed Central  Google Scholar 

  • Green MA, et al. [68Ga]Ga-P16-093 as a PSMA-targeted PET radiopharmaceutical for detection of cancer: initial evaluation and comparison with [68Ga]Ga-PSMA-11 in prostate cancer patients presenting with biochemical recurrence. Mol Imaging Biol. 2020;22(3):752–63.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • Heilinger J, et al. Threshold for defining PSMA-positivity prior to Lu-PSMA therapy: a comparison of [Ga]Ga-PSMA-11 and [F]F-DCFPyL in metastatic prostate cancer. EJNMMI Res. 2023;13:83.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • Houshmand S, Lawhn-Heath C, Behr S. PSMA PET imaging in the diagnosis and management of prostate cancer. Abdom Radiol (NY). 2023;48(12):3610–23.

    Article  PubMed  PubMed Central  Google Scholar 

  • Jani AB, et al. Diagnostic performance and safety of 18F-rhPSMA-7.3 positron emission tomography in men with suspected prostate cancer recurrence: results from a phase 3, prospective, multicenter study (SPOTLIGHT). J Urol. 2023;210(2):299–311.

    Article  PubMed  Google Scholar 

  • Kersemans K, et al. Automated radiosynthesis of Al[18F]PSMA-11 for large scale routine use. Appl Radiat Isot. 2018;135(May):19–27.

    Article  CAS  PubMed  Google Scholar 

  • Lee H, et al. Preliminary evaluation of 68Ga-P16-093, a PET radiotracer targeting prostate-specific membrane antigen in prostate cancer. Mol Imaging Biol. 2022;26:1–11.

    Google Scholar 

  • Link JM, Shoner SC, Krohn KA. Sources of carrier F-19 in F-18 fluoride. AIP Conf Proc. 2012;1509(1):61–5.

    Article  ADS  CAS  Google Scholar 

  • Malik N, et al. Radiofluorination of PSMA-HBED via AlF chelation and biological evaluations in vitro. Mol Imaging Biol. 2015;17:777–85.

    Article  MathSciNet  CAS  PubMed  Google Scholar 

  • McBride WJ, et al. A novel method of 18F radiolabeling for PET. J Nucl Med. 2009;50(6):991–8.

    Article  CAS  PubMed  Google Scholar 

  • Surasi DS, et al. Diagnostic performance and safety of positron emission tomography with 18F-rhPSMA-7.3 in patients with newly diagnosed unfavourable intermediate-to very-high-risk prostate cancer: results from a phase 3, prospective, multicentre study (LIGHTHOUSE). Eur Urol. 2023;84(4):361–70.

    CAS  Google Scholar 

  • Thisgaard H, et al. Multi-curie production of gallium-68 on a biomedical cyclotron and automated radiolabelling of PSMA-11 and DOTATATE. EJNMMI Radiopharm Chem. 2021;6(1):1.

    Article  PubMed  PubMed Central  Google Scholar 

  • Voter AF, et al. Piflufolastat F-18 (18F-DCFPyL) for PSMA PET imaging in prostate cancer. Expert Rev Anticancer Ther. 2022;22(7):681–94.

    Article  CAS  PubMed  Google Scholar 

  • Wang G, et al. A prospective head-to-head comparison of [(68)Ga]Ga-P16-093 and [(68)Ga]Ga-PSMA-11 PET/CT in patients with primary prostate cancer. Eur J Nucl Med Mol Imaging. 2023;50(10):3126–36.

    Article  CAS  PubMed  Google Scholar 

  • Wurzer A, et al. Radiohybrid ligands: a novel tracer concept exemplified by (18)F- or (68)Ga-labeled rhPSMA inhibitors. J Nucl Med. 2020;61(5):735–42.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  • Wurzer A, et al. Automated synthesis of [(18)F]Ga-rhPSMA-7/-7.3: results, quality control and experience from more than 200 routine productions. EJNMMI Radiopharm Chem. 2021;6(1):4.

    Article  PubMed  PubMed Central  Google Scholar 

  • Yao X, et al. Optimization of solid-phase extraction (SPE) in the preparation of [(18)F]D3FSP: a new PET imaging agent for mapping Aβ plaques. Nucl Med Biol. 2019;71:54–64.

    Article  CAS  PubMed  Google Scholar 

  • Zha Z, et al. Synthesis and evaluation of a novel urea-based 68Ga-complex for imaging PSMA binding in tumor. Nucl Med Biol. 2018;59:36–47.

    Article  CAS  PubMed  Google Scholar 

  • Zha Z, et al. Radiolabeling optimization and preclinical evaluation of the new PSMA imaging agent [(18)F]AlF-P16-093. Bioconjug Chem. 2021;32(5):1017–26.

    Article  CAS  PubMed  Google Scholar 

  • Zhao R, et al. VMAT2 imaging agent, D6-[18F]FP-(+)-DTBZ: improved radiosynthesis, purification by solid-phase extraction and characterization. Nucl Med Biol. 2019;72–73:26–35.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors thank Dr. Hsiaoju Lee, Dr. Alexander Schmitz, David Schaub and Lawrence Toto at the University of Pennsylvania cyclotron facility for providing [18F]fluoride. The authors thank Dr. William Eckelman for his helpful discussions and comments during preparation of this manuscript.

Funding

Funding for research was provided by Five Eleven Pharma.

Author information

Authors and Affiliations

Authors

Contributions

DA, SRC, KP, DK, RZ, LZ and HK contributed to the overall study concept and design. DA and DK were responsible for the design and construction of the custom automated systems (PiBOX) used for all experiments. DA, SRC and KP were responsible for material preparation, data collection and analysis. DA was responsible for the first draft of the manuscript and SRC, KP, DK, RZ, LZ and HK had the opportunity to comment on subsequent versions and have read and approved the final manuscript.

Corresponding author

Correspondence to David Alexoff.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

DA, SRC and KP are employees of Five Eleven Pharma. HK is the founder and Chairman of the board at Five Eleven Pharma. DK has received fees as a consultant for Five Eleven Pharma. Other authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Alexoff, D., Choi, S.R., Ploessl, K. et al. Optimization and scale up of production of the PSMA imaging agent [18F]AlF-P16-093 on a custom automated radiosynthesis platform. EJNMMI radiopharm. chem. 9, 15 (2024). https://doi.org/10.1186/s41181-024-00247-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41181-024-00247-1

Keywords