Skip to main content

Table 5 Radiochemical purity assessed using HPLC method during 4 h

From: Automated synthesis and quality control of [68Ga]Ga-PentixaFor using the Gaia/Luna Elysia-Raytest module for CXCR4 PET imaging

RCP using HPLC method

 

Batch 1

Batch 2

Batch 3

Final vial radioactivity

586 MBq

661 MBq

594 MBq

T = 0 h

RCP

99.70%

99.80%

100%

99.80%

99.80%

100%

99.80%

99.70%

99.90%

Free Ga3+

0.32%

0.23%

0.05%

0.18%

0.16%

0.05%

0.16%

0.32%

0.11%

T = 1 h

RCP

99.90%

99.80%

99.90%

99.80%

99.80%

99.90%

100%

99.80%

99.80%

Free Ga3+

0.15%

0.20%

0.06%

0.18%

0.19%

0.13%

0.03%

0.22%

0.16%

T = 2 h

RCP

99.90%

99.90%

99.80%

100%

99.90%

99.80%

100%

99.90%

99.90%

Free Ga3+

0.09%

0.09%

0.17%

0.05%

0.15%

0.20%

0.03%

0.12%

0.13%

T = 3 h

RCP

99.90%

99.80%

99.80%

99.90%

99.70%

99.90%

99.90%

99.80%

99.80%

Free Ga3+

0.10%

0.24%

0.17%

0.15%

0.30%

0.12%

0.14%

0.19%

0.23%

T = 4 h

RCP

100%

99.90%

99.90%

100%

99.70%

99.90%

100%

99.90%

99.70%

Free Ga3+

0.02%

0.13%

0.10%

0.05%

0.27%

0.15%

0.02%

0.06%

0.30%