Skip to main content

Table 1 Overview of the different samples for GC measurements and the activity at the start of the measurements

From: Gamma counting protocols for the accurate quantification of 225Ac and 213Bi without the need for a secular equilibrium between parent and gamma-emitting daughter

Sample

225Ac (in SEq with 213Bi)

Added 213Bi

Experiment

1

1.5 MBq (total)

0 kBq

GC calibration

2

0 kBq

117 kBq

Quantifying down scatter of higher-energy gamma emissions by 213Bi in \({\text{EW}}_{{221_{{{\text{Fr}}}} }}\)

3

94 kBq

55 kBq

Simulating non-SEq conditions between 225Ac and 213Bi

4

116 kBq

32 kBq

5

1–2 MBq

0 kBq

Radiolabeling of [225Ac]Ac-DOTA and incubation in human serum for 15 min and 24 h

6

1–2 MBq

0 kBq

Radiolabeling of [225Ac]Ac- DEPA and incubation in human serum for 15 min and 24 h

  1. Specifications of the samples used for determining the linear range of GC measurements and evaluating the impact of different sample volumes on GC measurements is given in the Additional file 1