Table 2 Distribution of radioactivity after automated-synthesis of [18F]AlF-FAPI-74 using CFN-MPS200

From: One-pot and one-step automated radio-synthesis of [18F]AlF-FAPI-74 using a multi purpose synthesizer: a proof-of-concept experiment

Measured point Radioactivity distribution (%)
Product vial ([18F]AlF-FAPI-74) 37.0 ± 4.3
18O-Water recovery vial 3.1 ± 1.4
QMA cartridge, 130 mg 2.3 ± 1.3
18O-Water waste liquid vial 0.7 ± 1.0
Reactor 8.6 ± 1.5
Dilution vial 0.5 ± 0.2
HLB cartridge 0.3 ± 0.1
Waste liquid vial 45.9 ± 4.0
Formulation vial 0.7 ± 0.4
Synthesis cassette 1.0 ± 0.3
  1. (n = 10)