Skip to main content

Table 1 Results of the optimisation study with tosylate DPA-714 precursor in anhydrous acetonitrile at 90C after 10 min with different PTC/base systems. Each run was performed with the indicated volume of the PTC/base combination dissolved in acetonitrile/water 70/30 v/v% for Entry 1 and 85/15 v/v% for Entries 2 and 3

From: Optimised GMP-compliant production of [18F]DPA-714 on the Trasis AllinOne module

PTC

Base

Elution volume [mL]

Radiochemical yield [%]*

[18F]fluoride [%]a

K222 (37.2 mM)

K2CO3 (11.0 mM)

1

77

23

K222 (30.0 mM)

KHCO3 (30.0 mM)

0.75

84

15

TEAB (40.0 mM)

-

0.6

94

5

  1. anon-isolated, estimated by (radio)HPLC