Nuclide (activity)

Measured Dose rate at 10 cm with 0.4 mm tungsten shielding (µSv/h)

Theoretical dose rate at 10 cm with 0.4 mm tungsten shielding (µSv/h)

Lu-177 (7400 MBq)

13.5

9.3

Zr-89 (37 MBq)

150

393

Ga-68 (150 MBq)

300

F-18 (300 MBq)

1200

1902