Skip to main content

Table 3 Overview of the bias and coefficient of variation for different syringe volumes

From: Validation of an automated dispensing system for subsequent dose dispensing of different radionuclides

Nuclide

Syringe volume (mL)

Nominal activity (MBq)

Bias (%)

(Criteria: <  ± 5%)

Coefficient of variation (%)

(Criteria: <  ± 5%)

177Lu

30

100

0.77

1.70

20

100

0.16

0.74

3

40

− 2.00

3.03

89Zr

30

37

− 0.30

0.62

20

37

− 0.38

0.39

3

5

2.35

1.70

68 Ga

10

500

− 0.25

− 0.59

5

300

− 0.19

0.36

3

100

0.15

0.43

18F

10

1,000

− 0.51

0.19

5

500

0.03

0.53

3

100

− 0.17

0.47