Table 2 Overview of the validated linearity parameters for each nuclide

From: Validation of an automated dispensing system for subsequent dose dispensing of different radionuclides

Nuclide

177Lu

89Zr

68Ga

18F

Validated range (MBq)

1000–10,000

20–74

10–1000

100–2000

Mean deviation in dispensed dose over three days (n = 3 per day) per nomimal activity

(Criteria: <  ± 5%)

− 0.46% (10,000 MBq)

− 0.29% (7400 MBq)

− 0.50% (5000 MBq)

− 0.34% (3700 MBq)

− 0.76% (1000 MBq)

− 0.15% (74 MBq)

− 0.27% (50 MBq)

− 0.20% (37 MBq)

− 0.30% (20 MBq)

3.57% (10 MBq)

− 0.70% (1000 MBq)

− 0.50% (500 MBq)

− 0.35% (150 MBq)

− 0.52% (100 MBq)

− 0.51% (50 MBq)

− 1.15% (10 MBq)

− 0.26% (2000 MBq)

− 0.24% (1500 MBq)

− 0.36% (1000 MBq)

− 0.28% (500 MBq)

− 0.49% (100 MBq)

Mean r2 calibration curve (n = 3)

(Criteria: > 0.990)

1.000

0.9999

1.000

1.000

95% confidence interval intercept

(Criteria: interval contains zero)

Complies for each run (n = 3)

Complies for each run (n = 3)

Complies for each run (n = 3)