Skip to main content

Table 1 Measuring range of the automated dispensing system (ADS) for various nuclides

From: Validation of an automated dispensing system for subsequent dose dispensing of different radionuclides

Nuclide

Accurate operational range as specified by manufacturer (MBq)

Range for validation based on locally used activity doses (MBq)

Nominal activity concentration in multidose vial

177Lu

100–10,000

1000–10,000

1000 MBq/mL

89Zr

0.2–50

10–74

10 MBq/mL

68Ga

10–2000

10–1000

100 MBq/mL

18F

10–3000

100–2000

185 MBq/mL