Skip to main content

Table 6 Acceptance limits and parameters for the validation of analytical methods (to assure that the analytical methods can properly determine specifications set in Table 2)

From: Optimized, automated and cGMP-compliant synthesis of the HER2 targeting [68Ga]Ga-ABY-025 tracer

HPLC

Specificity

Sample matrix spiked with ABY-025: Area of ABY-025 peak in spiked matrix differ maximum ± 20% and that no peaks ≥ 10% of the area of 50 µg/mL

ABY-025 in the range 6–10 min in UV

Sample matrix spiked with 68Ga: 68Ga should elute only in the void peak in the RD

Difference in Rt between [68Ga]Ga-ABY-025 (radio chromatogram) and Rt ABY-025 (UV-chromatogram of SST) ≤ 60 s

Linearity of chemical purity for ABY-025

Peak area in at least 5 levels in the range 30–100 µg/mL. Linear regression should give a r2 ≥ 0.99

Carry over: The ABY-025 peak in blank after the 100 µg/mL sample should not exceed 20% of peak in previous injection

Linearity of radiochemical purity for [68Ga]Ga-ABY-025

Linearity is tested for samples from 10 MBq/mL to 150% of normal product activity (75 MBq/mL)

Linear range for the method are defined as the activity range where %ROI [68Ga]Ga-ABY-025 differs ≤ 0.5%ROI for the average (of samples within range)

Repeatability

The repeatability is tested in the range 30–100 µg/mL ABY-025. The relative response, as compared with the SST should show an RSD% ≤ 10% on each concentration level (3 injections) and for all injections (9 injections) within the range 30–100 µg/mL (3 concentrations / 3 replicates). Relative response = (Area/Conc.)Sample / (Area/Conc.)SST/Ref

Intermediate precision

On 3 occasions 3 triplicates on 3 concentrations are performed. Different days or analysts, at least two columns from different lots. The relative response, as compared with the SST should show an RSD% ≤ 10% on each concentration level (9 injections) and for each analysist (9 injections)

Accuracy

Accuracy can be inferred once precision, linearity and specificity has been established. Accuracy should be shown in the range 30–100 µg/mL

GC

Linearity

R2 ≥ 0.99; Slope, y-intercept and residual sum of square must be given. A minimum of 5 concentrations must be used

LOD and LOD: Calculation based the standard deviation of the response and the slope (based on calibration curve)

Repeatability

RSD% ≤ 10% for a minimum of 9 determinations over 3 concentrations within the range (e.g. 3 concentrations /3 replicates from Accuracy)

Intermediate precision

RSD% ≤ 10%; for a minimum of 9 determinations over 3 concentrations within the range (e.g. 3 concentrations/3 replicates from Accuracy) performed a different day or by a different analyst

Accuracy

Diff% ≤ 10% between calculated and actual concentrations for 3 concentrations within the range using a minimum of 3 replicate determinations from each concentration

TLC

Specificity

i.e. its ability to determine free 68Ga and/or unidentified radioactive impurities. The following criteria must be met:

• The peak corresponding to free 68Ga and/or unidentified radioactive impurities must be completely separated from the peak corresponding to [68Ga]Ga-ABY-025

• The radioactivity of free 68Ga and/or unidentified radioactive impurities must be ≤ 3% of the total radioactivity to be released

• Retardation factors (Rf) ~ 0–0.1 for free 68Ga and/or unidentified radioactive impurities Rf ~ 0.2–0.6 and Rf > 0.6 for [68Ga]Ga-ABY-025

Endotoxins

Optimal dilution factor

Interference screen on worst case simulation of [68Ga]Ga-ABY-025 formulation using maximum allowed ethanol concentration

The optimal dilution factor is the one that yields a spike recovery close 100% (related to the known endotoxin level in the cartridge)

Sterility

MST

MST at the external contractor according to their certified protocol

  1. RD = Radiodetector, ROI = Region of interest, Rt = Retention time, SST = Suitability test