Skip to main content

Table 1 Radiosynthesis results

From: Fully automated radiosynthesis of [68Ga]Ga-FAPI-46 with cyclotron produced gallium

Quality control test

Specifications

Qual Run 1

Qual Run 2

Qual Run 3

Appearance

Clear, colorless, no particulates

Clear, colorless, no particulates

Clear, colorless, no particulates

Clear, colorless, no particulates

Filter integrity

Pass

Pass

Pass

Pass

pH

4.0–7.0

4.5

4.5

4.5

Radiochemical purity (%) (HPLC)

NLT 90% *

95.31

95.24

96.40

Radiochemical purity (%) (iTLC)

NLT 90%

> 99

> 99

> 99

Radionuclidic identity (t1/2, minutes)

64–67

64.46

67.72

69.65

Bacterial endotoxin levels (EU/mL)

< 8.75 EU/mL

< 1.00 EU/mL

< 1.00 EU/mL

< 1.00 EU/mL

Sterility (observed growth after 14 d)

No growth

No growth

No growth

No growth

Radiochemical identity (% difference from reference Std) by HPLC

NMT 10% *

1.46%

1.55%

1.30%

Amount of product @ EOS

N/A

45.8 mCi (1695 MBq)

41.1 mCi (1521 MBq)

44.2 mCi (1635 MBq)

Gallium-66 and Gallium-67 impurities @ EOS

< 2.0%*

0.132%

0.145%

0.159%

  1. NLT not less than, NMT not more than, EOS end of synthesis
  2. *Periodic quality indicator test