Skip to main content

Table 7 Results of [68Ga]Ga-EMP100 stability study in final vial

From: Fully automated radiolabeling of [68Ga]Ga-EMP100 targeting c-MET for PET-CT clinical imaging

Test

Method

Specifications

Batch 1

Batch 2

Batch 3

Aspect

Visual

Clear, colourless solution

Yes

Yes

Yes

pH value

pH-strips

4.0–8.0

6.0

6.0

6.0

Stability over 3 h (RCP %)

Radio HPLC

[68Ga]Ga-EMP100 > 95,00%

98.63%

98.87%

99.52%

  

Free [68Ga]GaCl3 ≤ 5%

1.37%

1.13%

0.48%

 

Radio TLC AcNH4 1 M Methanol (V/V)

[68Ga]Ga-EMP100 > 95%

99.85%

99%

99.91%

  

[68Ga]Ga-colloidal ≤ 5%

0. 15%

1.00%

0.09%

 

RCP (overall)

[68Ga]Ga-EMP100 (overall) > 95.00%

99.08%

97.87%

99.43%

Sterility

Inoculation 14 days (Ph Eur 20601)

Sterile

Sterile

Sterile

Sterile