Skip to main content

Table 2 Outline of the [18F]DPA-714 cassette on the Trasis AllinOne radiosynthesiser, with description of reagent and materials position. Positions outlined in boldface are original parts of the commercially available [18F]FDOPA cassette

From: Optimised GMP-compliant production of [18F]DPA-714 on the Trasis AllinOne module

Manifold position

Reagents or materials attached

1 vertical

PE tubing to [18O]H2O recovery vial

1 horizontal

Silicone tubing to pressure inlet

2

Spike, white Luer Lock ring and barrel for 20 mm vial

3

3 mL sterile syringe with Luer Lock

4

Silicone tubing connected to QMA cartidge in position 5

5

Activated QMA cartridge

6

3 mL sterile syringe with Luer Lock (activity inlet)

7

Reactor 6 mL with 2 tubes (male connector to cassette, female to exhaust)

8

Spike, red Luer Lock ring with barrel for 20 mm vial

9

10 mL BD PlastiPak sterile syringe with Luer Lock

10

Spike, blue Luer Lock ring

11

PE tubing OD 2 mm length 150 mm, connected to Load port of HPLC unit

12

Spike, green Luer Lock ring with barrel for 20 mm vial

13

PE tubing OD 2 mm length 150 mm, connected to 25 mL HPLC product receiving vial via sterile needle

14

20 mL HSW Norm-Ject sterile syringe with Luer Lock

15

PE tubing OD 2 mm length 100 mm, connected to Sep-Pak C18 Plus cartridge in position 17

16

0.9% NaCl for injection bottle, connected to valve using PE tubing OD 2 mm length 150 mm with spike

17

Activated Sep-Pak C18 Plus cartridge

18 vertical

Tubing to dispensing hot cell

18 horizontal

Silicone tubing to exhaust

Other components

Reagents or materials attached

Collect port HPLC unit

PE tubing OD 2 mm length 150 mm, connected to 25 mL HPLC product receiving vial via sterile needle

HPLC product collection 25 mL vial

Connected to cassette via tubing in Load port and postion 13

QMA reagent for elution (10 mL vial)

Connected to position 2 via spike

Precursor solution (10 mL vial)

Connected to position 8 via spike

Quench solution (25 mL vial)

Connected to position 10 via spike

Ethanol solution (10 mL vial)

Connected to position 12 via spike

0.9% NaCl for injection solution (250 mL bottle)

Connected to position 16 via tubing and spike