Table 6 Comparison downscaled conditions to clinical conditions (t = 0 h)

From: Radiolabeling and quality control of therapeutic radiopharmaceuticals: optimization, clinical implementation and comparison of radio-TLC/HPLC analysis, demonstrated by [177Lu]Lu-PSMA

 

RCP (t = 0, %)

74 MBq

96.2 ± 1.0 (n = 3)

740 MBq

98.7 ± 0.5 (n = 3)

18 GBq

97.3 ± 0.8 (n = 3)

34 GBq

97.1 ± 0.5 (n = 3)