Skip to main content

Table 5 Robustness of the counter

From: Validation of a cost-effective alternative for a radiochromatography method to be used in a developing country

 

%RCP Mean ± RSD

%RCP Mean ± RSD

%RCP Mean ± RSD

Change

counter at normal position for both parts

counter tilted 5° for the strip section containing the impurity

counter tilted 5° for the strip section with the bound compound

MIBI (n = 5)

91.5 ± 1.0

91.8 ± 2.6 (P = 2.9*)

91.9 ± 2.3 (P = 2.4*)

  1. *P values for ANOVA of the respective RCP compared to RCP obtained with the counter in the normal position