Table 6 Results of impurities found applying deviations in solvent mixtures for strip 1

From: A new two-strip TLC method for the quality control of technetium-99m mercaptoacetyl-triglycine (99mTc-MAG3)

Sample Strip 1: Ratio EtAc/MEK
55:45 60:40 65:35
Impurities (%) Impurities (%) Impurities (%)
Sample 1 6.5 6.1 6.0
Sample 2 6.6 6.5 6.0
Sample 3 6.7 6.4 6.6
Average 6.6 6.3 6.2
STD (k=2) 0.2 0.4 0.6