Table 5 Comparison of RCP values determined by the new TLC method by scanner or dose calibrator against the RCP obtained by the Ph.Eur. method (HPLC and PC) taken as reference (standard deviation k=2)

From: A new two-strip TLC method for the quality control of technetium-99m mercaptoacetyl-triglycine (99mTc-MAG3)

   Ph.Eur. 2-strip method Accuracy
Date Validation (HPLC&PC) TLC Scanner Dose calibrator Difference to Ph.Eur. method
   [%] [%] [%] TLC Scanner Dose calibrator
      [%] [%]
27.04.2016 No 1 97 96.7 ± 0.8 97.1 ± 0.8 -0.3 0.1
28.04.2016 No 2 96.9 96.7 ± 0.5 97.1 ± 0.8 -0.2 0.2
29.04.2016 No 3 95.8 96.1 ± 0.4 Not measured 0.3 N.A.
29.04.2016 No 4 95.6 ± 0.5 96.1 ± 0.5 -0.2 0.3