|  | Ph.Eur. | 2-strip method | Accuracy | ||||
---|---|---|---|---|---|---|---|---|
Date | Validation | (HPLC&PC) | TLC Scanner | Dose calibrator | Difference to Ph.Eur. method | |||
 |  | [%] | [%] | [%] | TLC Scanner | Dose calibrator | ||
 |  |  |  |  | [%] | [%] | ||
27.04.2016 | No 1 | 97 | 96.7 | ± 0.8 | 97.1 | ± 0.8 | -0.3 | 0.1 |
28.04.2016 | No 2 | 96.9 | 96.7 | ± 0.5 | 97.1 | ± 0.8 | -0.2 | 0.2 |
29.04.2016 | No 3 | 95.8 | 96.1 | ± 0.4 | Not measured | 0.3 | N.A. | |
29.04.2016 | No 4 | 95.6 | ± 0.5 | 96.1 | ± 0.5 | -0.2 | 0.3 |