Table 5 Comparison of RCP values determined by the new TLC method by scanner or dose calibrator against the RCP obtained by the Ph.Eur. method (HPLC and PC) taken as reference (standard deviation k=2)

From: A new two-strip TLC method for the quality control of technetium-99m mercaptoacetyl-triglycine (99mTc-MAG3)

  

Ph.Eur.

2-strip method

Accuracy

Date

Validation

(HPLC&PC)

TLC Scanner

Dose calibrator

Difference to Ph.Eur. method

  

[%]

[%]

[%]

TLC Scanner

Dose calibrator

     

[%]

[%]

27.04.2016

No 1

97

96.7

± 0.8

97.1

± 0.8

-0.3

0.1

28.04.2016

No 2

96.9

96.7

± 0.5

97.1

± 0.8

-0.2

0.2

29.04.2016

No 3

95.8

96.1

± 0.4

Not measured

0.3

N.A.

29.04.2016

No 4

95.6

± 0.5

96.1

± 0.5

-0.2

0.3