A new automated and putatively versatile synthesis of the PSMA-ligand derivative [18F]DCFPyL using the FASTlabTM synthesizer

Abstract

Background

Noninvasive molecular imaging using peptides and biomolecules labelled with positron emitters has become important for detection of cancer and other diseases with PET (positron emission tomography). The positron emitting radionuclide fluorine-18 is widely available in high yield from cyclotrons and has favorable decay (t1/2 109.7 min) and imaging properties. 18F-Labelling of biomolecules and peptides for use as radiotracers is customarily achieved in a two-step approach, which can be challenging to automate. 6-[18F]Fluoronicotinic acid 2,3,5,6-tetrafluorophenyl ester ([18F]F-Py-TFP) is a versatile 18F-prosthetic group for this purpose, which can be rapidly be produced in an one-step approach on solid support. This work details an automated procedure on the cassette-based GE FASTlab™ platform for the labeling of a peptidomimetic, exemplified by the case of using the Glu-CO-Lys motif to produce [18F]DCFPyL, a ligand targeting the prostate specific membrane antigen (PSMA).

Results

From fluorine-18 delivery a fully automated two-step radiosynthesis of [18F]DCFPyL was completed in 56 min with an overall end of synthesis yield as high as 37% using solid phase extraction (SPE) purification on the GE FASTlab™ platform.

Conclusions

Putatively, this radiolabeling methodology is inherently amenable to automation with a diverse set of synthesis modules, and it should generalize for production of a broad spectrum of biomolecule-based radiotracers for use in PET imaging.

Background

Cancer is the leading cause of premature deaths in the western world, a circumstance that has spurred the development of new radiopharmaceuticals for cancer diagnosis by positron emission tomography (PET) and for radiotherapy using α- and β-emitting radionuclides (Bouchelouche et al. 2010; Holland et al. 2012; Zhang et al. 2017). For use in clinical PET imaging, the radiohalogen fluorine-18 (18F) is the preferred radioisotope due to its physical half-life (109.7 min), favorable decay properties, and its ready availability in multi-GBq quantities from medical cyclotrons. A considerable proportion of emerging radiopharmaceuticals are based on small to medium sized peptides, fusion proteins, and endogenous proteins. However, these products are not always obtainable through a direct radiolabeling approach due to their instability at high temperature or the otherwise harsh reaction conditions required to introduce fluorine-18 into their structure. Although labeling of such molecules with fluorine-18 using a chelator conjugated to the targeting moiety with [18F]AlF (aluminum [18F]fluoride) is feasible, this approach can release free [18F]fluoride in vivo (which invariably results in intense bone labelling) and otherwise alter the biodistribution properties of the radiopharmaceutical in an unpredictable fashion (Lütje et al. 2019).

A robust and high yield radiosynthesis amenable to automation is a prerequisite for adopting a PET radiopharmaceutical for widespread clinical or research use (Li et al. 2013). Production of PET radiopharmaceuticals with a multi-step synthetic process can be a challenging task using commercially produced cassette-based modules, mainly due to their lack of modifiability, and the frequent requirement for intermediate or final HPLC purification. 18F-labelled tracers for prostate specific membrane antigen (PSMA) [18F]DCFPyL and [18F]-PSMA-1007 are examples of PET radiopharmaceuticals that formerly required a two-step radiosynthesis from the prosthetic group 6‐[18F]fluoronicotinic acid‐2,3,5,6‐tetrafluorophenyl ester ([18F]F‐Py‐TFP) (Chen et al. 2011; Giesel et al. 2017, Cardinale et al. 2017). In the case of [18F]DCFPyL, the first reported radiosynthesis from [18F]F‐Py‐TFP and the PSMA-avid substrate Glu-CO-Lys protected with t-butyl has subsequently been optimized (Chen et al. 2011). Currently, a direct precursor labeling with the trimethylammonium nicotinic acid moiety conjugated to the ε-nitrogen of the lysine residue of Glu-CO-Lys is the favored method for automated multi-dose production of [18F]DCFPyL (Dornan et al. 2018), which has greatly facilitated the use of PSMA PET in clinical routine (Fig. 1). However, this introduction of fluoride-18 in organic solvents at a temperature approaching 100 °C, is unlikely to provide high labeling efficiencies, or to generalize for larger molecular weight radiopharmaceuticals such as polypeptides and proteins with their bountiful functional groups.

Fig. 1
figure 1

Three Glu-CO-Lys-based PET radiopharmaceuticals presently in clinical use for PSMA imaging

Fairly recently, a simplistic and rapid preparation of [18F]F‐Py‐TFP has been reported, wherein the prosthetic group is obtained directly by passing its precursor in a solution of acetonitrile/tert-butanol through a polymeric anion‐exchange cartridge preloaded with [18F]fluoride (Basuli et al. 2016; Olberg and Svadeberg 2017). This procedure yields the bifunctional labeling agent in good yield, with an overall synthesis time of less than ten minutes (EOB), simultaneously avoiding the use of a reaction vessel and a phase transfer catalyst. The on-column radiosynthesis of [18F]F‐Py‐TFP has a very small laboratory footprint and lends itself well for automation and offer radiolabeling of a wide variety of sensitive biomolecules (proteins eg.) (Roy et al. 2019; Lesniak et al. 2019). Expanding on this concept, we report a fully automated, two-step radiosynthesis of [18F]DCFPyL from [18F]F‐Py-TFP and the Glu-CO-Lys pseudo-peptide precursor, which can serve as a blueprint manufacturing process for the preparation of other 18F-labelled radiopharmaceuticals. The approach uses [18F]F‐Py‐TFP produced on column adopted into the commercially available, cassette-based synthesis module FASTlab™. The automated process implements a hydrophilic interaction liquid chromatography (HILIC) SPE cartridge purification methodology, making final preparative HPLC purification redundant.

Methods

General

Chemicals were obtained either from Sigma-Aldrich (Norway) or from VWR (Norway), which were used without further purification. HPLC grade water was obtained from the Direct-Q® (EMD Millipore) water purification system. Reference standard 6-fluoronicotinic acid 2,3,5,6-tetrafluorophenyl ester (1) and its trimethylammonium precursor (2) were synthesized according to literature methods (Olberg et al. 2010). DCFPyL (2-(3-{1-carboxy-5-[(6-fluoro-pyridine-3-carbonyl)-amino]-pentyl}-ureido)-pentanedioic acid) (3) reference standard was purchased from ABX GmbH (Radeberg, Germany). Glu-CO-Lys (2-[3-(5-Amino-1-carboxy-pentyl)-ureido]-pentanedioic acid trifluoroacetate salt) (4) was purchased from ABX GmbH or Scintomics GmbH (Fürstenfeldbruck, Germany). Sep-Pak™ cartridges were purchased from Waters (Norway). HILIC cartridges containing 350 mg of Chromabond® HILIC sorbent (Macherey–Nagel GmbH, Düren) and anion-exchange columns containing 45 mg of Chromabond® PS-HCO3, (Macherey–Nagel GmbH) were obtained from Teknolab AS (Ski, Norway). Analytical HPLC was performed on an Agilent 1100 series HPLC system with UV detection connected in series with a NaI γ-radiodetector (Institute for Energy Technology (IFE), Kjeller, Norway). Samples were analyzed using methods 1 or 2a and b as described below. Chromatograms were processed with Agilent OpenLAB CDS EZChrom version A.04.04. Reversed phased-analytical HPLC was performed on a Phenomenex kinetex EVO C18 column (50 × 2.1 mm, 2.6 µm particles), at a flow rate of 0.5 mL/min, with a composition gradient from 2–95% solvent B (method 1) over ten min (solvent A: water/0.05% TFA, solvent B: acetonitrile/0.05% TFA). UV detection was at 214, 254 and 264 nm. Hydrophilic interaction liquid chromatography (HILIC) was performed on a SeQuant ZIC-cHILIC (100 × 4.6 mm, 3 µm particles) at a flow rate of 2.0 mL/min, with composition gradient from 2–98% solvent B (method 2a) or isocratically with 68% solvent B (method 2b) over ten min (Solvent A: 97% acetonitrile/3% 5 mM NH4OAc, solvent B: 5 mM NH4OAc). UV detection was at 210, 220, 254 and 264 nm. Thin layer chromatography (TLC) was run on gel 60F254 plates (Merck) using acetonitrile as eluent. A Raytest miniGita (Raytest, Germany) equipped with a β-detector was used to record the radio-TLC scan. Gas chromatography (GC) was performed with an Agilent 6890 N GC (Matriks, Norway) equipped with a flame ionization detector (FID), a column oven, and an auto-sampler with a direct injection system. The GC column was a fused silica capillary column with USP stationary phase G43 (6% cyanopropylphenyl—94% dimethyl polysiloxane) measuring 0.32 mm i.d. × 30 m. Automated radiochemistry was performed on the GE FASTlab synthesizer with the custom-made single-use FASTlab development kit (GE Healthcare) placed inside a lead-shielded hot cell. Fluorine-18 was produced in a GE PETtrace 6 cyclotron (General Electric Healthcare) with a GE 18F Nb 27 self-shielded target either by irradiation (40 µAh × two min) of [18O]H2O (Taiyo Nippon Sanso, Japan), or from a target rinse after a routine 100 GBq production. No more than 150 MBq of radioactivity were handled outside the hot cell. Radioactivity was measured with a Capintec dose calibrator (New Jersey, USA). The identities of the radiolabeled products were confirmed by co-injection to HPLC along with authentic nonradioactive standards.

General radiolabeling procedure (manual)

Aqueous [18F]fluoride was trapped on a Chromabond PS-HCO3 anion cartridge followed by rinsing with hot acetonitrile (2 mL) and passage of air (2 × 5 mL) to dryness. A solution of precursor 2 (10 mg/mL, 22 mM), in the presence of TEA in tert-butanol (t-BuOH)/acetonitrile (50%, v/v, 1 mL) was used to initiate radiolabeling and to elute the radioactivity into a receiving vial (10 s); the eluent was passed through the column back and forth into the syringe barrel for 20 s. After the final elution into the vial, the reaction was quenched by rinsing the column with water (0.5 mL). We measured the radioactivity concentration in the quenched reaction mixture before analysis by HPLC and TLC with on-line gamma detection.

General automated radiolabeling procedure

Cassettes were assembled according to the layout depicted in Fig. 2. External vials were vented with a syringe filter and connected to the cassette manifold with a silicone tube of specific length, terminating in a needle as follows:

  • 30 mL vial of H3PO4 0.85%: 42-cm long tube, 0.80 i.d. × 80 mm needle (vial 4)

  • Collection vial (10 mL) 13-cm long tube, 0.80 i.d. × 80 mm needle (vial 13)

  • 25% ethanol vial: 13-cm long tube, 2.10 mm i.d. × 80 mm needle (vial 12)

Fig. 2
figure 2

Layout of the [18F]DCFPyL FASTlab cassette. 1. Precursor vial 2. Chromabond PS-HCO3 cartridge 3. Sep-Pak tC18 plus light cartridge 4. 0.85% H3PO4 external vial 5. Glu-CO-Lys vial 6. Acetonitrile 7. Acetone 8. Water bag 9. Abs. ethanol 10. HILIC SPE cartridge 11. Oasis PRiME HLB plus light cartridge 12. 25% ethanolic solution external vial 13. Product collection vial. S1, S2 and S3, syringes 1, 2 and 3

Vials were prepared as follows:

  • Precursor vial: [18F]F-Py-TFP precursor 2 (15 mg, 31.4 µmol) in t-BuOH/acetonitrile (1.4 mL) and triethylamine (TEA) (0.4 eq.) (vial 1)

  • Glu-CO-lys vial: Glu-CO-lys (13.2 mg, 29.1 µmol), TEA-HCO3 (4.3 eq.) in 0.63 mL DMSO (vial 5)

  • Acetonitrile vial: 4.0 mL (vial 6)

  • Acetone and absolute ethanol vials: 4.6 mL each (vial 7 and vial 9)

Results

From the get-go, reaction conditions were investigated manually, aiming to determine suitable and robust conditions for on-cartridge synthesis of [18F]F-Py-TFP for translation to the automated synthesis. In brief, fluorine-18 was trapped on the Chromabond PS-HCO3 and then the column was dried with anhydrous acetonitrile. Fluorination occurred on-cartridge when the mixture of precursor 2 and base in acetonitrile/t-BuOH was cycled through the cartridge. The radiochemical yield is dependent of precursor concentration and the presence of an non-nucleophilic organic base helps to increase yield and reduce variantion between synthesis runs.

RCYs improved when the Chromabond column was dried using acetonitrile heated to 80 °C in the reactor vessel. The best RCY (78 ± 9%) was obtained using 11 mM of precursor and 0.2 equivalents of triethylamin (TEA) as the base. Translating these conditions (including SPE purification) into the FASTlab, [18F]F-Py-TFP was obtained with RCY of 63 ± 4% with TEA as the base. In our hands, yields were 15% lower using the automated FASTlab synthesis compared to manual synthesis and might be due to the longer fluid paths in the automated synthesis. For the automated FASTlab [18F]DCFPyL synthesis, 22 mM of precursor 2 and 8 mM of TEA in 1.0 mL acetonitrile/t-BuOH (50/50 v/v) was used routinly throug this work.

The automated radiosynthesis entailed three sequential processes, i.e. a ten min pre-labelling step, a 25 min on-cartridge radiolabeling step for [18F]F-Py-TFP (including purification), and a 31 min conjugation and purification step affording the final product [18F]DCFPyL (56 min in total). Due to the inherent dead volumes, when implementing the radiochemistry to the FASTlab synthesizer, the reagent vials were filled with slightly larger volumes to retain the same stoichiometric amounts of reagents.

Pre-labelling

The vials, cartridges and reactor were conditioned for ten minutes before delivery of radioactivity. Vials were pressurized with 1 bar nitrogen to ensure optimal syringe filling. Ethanol and water were successively used to condition the tC18 Sep-Pak light cartridge. Acetonitrile (1.6 mL) was transferred into the FASTlab reactor and heated; the system was deemed ready to receive radioactivity when the reactor temperature reached 80 °C.

Radiolabeling and purification of [18F]F-Py-TFP

The next step provided purified [18F]F-Py-TFP bound to the tC18 Sep-Pak light cartridge in 25 min. First, fluorine-18 was trapped in the Chromabond PS-HCO3 cartridge, followed by drying with two passages of hot acetonitrile drawn from the heated reactor. The acetonitrile was successively sent to syringes 2 and 3 to remove any water remaining from the pre-labeling step. Using syringe 1, a 1 mL volume of the precursor solution was pushed through the Chromabond PS-HCO3 cartridge over a period of 45 s, with a series of slow back and forth strokes. The radioactive mixture was then transferred into syringe 2 and quenched by rinsing of the Chromabond PS-HCO3 with water. Syringe 2 was further loaded with two additional aqueous rinses from the Chromabond PS-HCO3 cartridge. Radiochemically pure [18F]F-Py-TFP was obtained by diluting the organic solvent content with water to 20% and extracting the [18F]F-Py-TFP onto the tC18 Sep-Pak light cartridge (flow 2 mL/min), which was followed by a 6 mL water rinse of the tC18 via syringe 2. Partially dry SPE-bound [18F]F-Py-TFP was obtained by purging the SPE cartridge with a stream of nitrogen gas (N2) at high flow rate. Figure 3 and Table 1 present an overview of the cassette synthesis and the impurities removed by each of the SPE cartridges used in the FASTlab cassette.

Fig. 3
figure 3

Overview of the radiosynthesis process on the FASTlab

Table 1 Overview of impurities removed at each SPE-purification step

Reaction of [18F]F-Py-TFP to Glu-CO-Lys and purification of [18F]DCFPyL

From cartridge-bound [18F]F-Py-TFP, [18F]DCFPyL was synthesized and purified in 31 min.

The reactor was charged with Glu-CO-Lys/TBA-HCO3 in DMSO by applying vacuum, and the [18F]F-Py-TFP was then quantitatively eluted from the tC18 Sep-Pak light cartridge into the reactor using 0.7 mL absolute ethanol. Aminolysis of [18F]F-Py-TFP with Glu-CO-Lys was performed at 45 °C in the presence of TBA-HCO3 as base for 10 min, providing crude [18F]DCFPyL. In parallel, the cassette manifold was prepared for the final purification step. The HILIC and Oasis PRiME HLB plus light cartridges were both first conditioned with water from syringe 3. Syringe 2 was first rinsed with ethanol; (1 mL) and then acetone (1.2 mL), which was used to condition the HILIC. After conjugation, the reaction mixture was diluted with acetone (2.4 mL) and transferred via syringe 2 to the HILIC cartridge, immobilizing the [18F]DCFPyL product while more apolar impurities were unretained. Residual radioactivity was transferred from the reactor to the HILIC cartridge using a second portion of acetone (1.0 mL). The remaining acetone in vial 7 (approx. 1 mL) was used to rinse the HILIC cartridge followed by a N2-gas purge. Syringe 2 was rinsed with 0.85% phosphoric acid, and a 4.2 mL volume of this same acid solution was used to elute the [18F]DCFPyL fraction off the HILIC column to syringe 3, which was further diluted with 2.7 mL of 0.85% phosphoric acid.

Considering that both [18F]DCFPyL and unreacted precursor Glu-CO-Lys were retained by the HILIC SPE and both was eluted with weak phosphoric acid, an additional purification step to remove unreacted Glu-CO-Lys, which is a pseudo-carrier possessing PSMA receptor avidity, was required. After investigating different reverse solid-phase materials, including C18 and other polymeric phases, we identified the Oasis PRiME HLB plus light cartridge as being suitable, by retaining [18F]DCFPyL but not the more polar Glu-CO-Lys. Consequently, [18F]DCFPyL could be trapped selectively on the Oasis PRiME HLB light cartridge using syringe 2 as a waste receiver. The cartridge was washed with an additional 4.8 mL of phosphoric acid to elute any residual Glu-CO-Lys. Finally, syringe 3 was washed three times with water and [18F]DCFPyL was eluted off the Oasis PRiME cartridge into the product vial with 2.1 mL 25% ethanolic solution.

Quality control of the drug product

HPLC method 1 (reversed phase) was used in conjunction with on-line radioactivity detection for routine assessment of radiochemical and chemical purity. Chemical and radiochemical purity (RCP) of the undiluted product was also monitored using method 2a. The HILIC HPLC method 2b was used for quantifying in the purified product any residual Glu-CO-Lys, which was too hydrophilic to be resolved by reversed phased HPLC. Representative chromatograms from individual steps of the automated synthesis are depicted in Fig. 4. The results for four automated runs, using starting activities ranging from 1 to 95 GBq along with low activity automated runs (≤ 400 MBq), are shown in Table 2. The radiochemical yield was close to 40% irrespective of the starting activity using starting activities from 1 GBq and upwards. Furthermore, the radiochemical purity was never lower than 93% (range 93–98%) at end of synthesis (Table 3), and declined in the first hours after high radioactivity productions, in agreement with already published work showing radiolysis of multi-GBq productions of [18F]DCFPyL (Ravert et al. 2016). Dilution and final formulation of the drug product to reduce RAC and inclusion of a scavenger such as sodium ascorbate, is expected to stabilize RCP further.

Fig. 4
figure 4

HPLC chromatograms at separate stages of the radiosynthesis A HPLC analysis of [18F]F-Py-TFP automated on the FASTlab synthesizer. Solid red line; radiodetector (Method 1). B Reaction mixture after conjugation of [18F]F-Py-TFP with Glu-CO-Lys (4) prior to SPE cartridge purification from the FASTlab synthesizer. Solid red line; radiodetector. Solid blue line; UV detection at 254 nm. C HPLC analysis of SPE cartridge purified [18F]DCFPyL from a fully automated radiosynthesis using the FASTlab synthesizer. Solid red line; radiodetector. Solid blue line; UV detection at 254 nm

Table 2 FASTlab process yield of the automated synthesis of [18F]DCFPyL
Table 3 Main quality control parameters at different starting activities

Discussion

In the final radiopharmaceutical product, carrier DCFPyL is the main constituent of the PSMA-avid impurities; its concentration increases with higher starting radioactivities. The total chemical impurities were well below 200 µg per batch, such that dilution and formulation to a final 20 mL volume drug product would give concentration of less than 10 µg/mL. This is below the chemical purity specifications adopted for PSMA-1007 (sum of all impurities bellow 0.5 mg/Vmax and 0.1 mg/Vmax for carrier PSMA-1007) (Cardinale et al. 2017). Although the chemical impurities in this work is higher than in [18F]DCFPyL productions where final purification was performed with HPLC (20 nmol/mL [sum of all impurities] vs. a maximum of 0.88 nmol/mL [sum all unknown impurities for HPLC-purified [18F]DCFPyL]) (Windhorst 2019). The overall molar activity was moderately high, increasing sevenfold as starting radioactivity increased from 1 to 95 GBq. Nevertheless, the entire range of obtained molar activity is suitable for PET imaging of highly expressed receptor targets such as PSMA (7 GBq/µmol) (Pillarsetty et al. 2016). Residual solvents in the final drug product (acetone, TEA, acetonitrile, DMSO and t-BuOH) where quantified using GC-FID and were found to be well below the Class II (acetonitrile) and Class III residual solvent limits specified in the European Pharmacopeia.

Conclusion

We developed a fully automated two-step radiosynthesis of the PSMA ligand [18F]DCFPyL employing the prosthetic group [18F]-Py-TFP produced on-cartridge followed by conjugation to the PSMA-binding moiety Glu-CO-Lys. The radiosynhesis omits a preparative HPLC purification step. Our adoption of this reaction in the FASTlab™ synthesiser affords convenient access to the PSMA tracer [18F]DCFPyL in radiochemical yields ranging from 25 to 37% (uncorrected). The process, using GMP compliant single-use cassettes, afforded [18F]DCFPyL in adequate radiochemical purity (93–98%) and molar activity (70 ± 47 GBq/µmol), with radiochemical yields sufficient to support multiple clinical PET investigations from a single production. Residual solvents were well below the concentration limits adopted in the current version of the European Pharmacopeia. Furthermore, the fully automated proof-of-concept two-step process bodes well for a more general application to other radiosynthesis modules and for conjugation with large biomolecules such as oligopeptides and proteins.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

PSMA:

Prostate specific membrane antigen

[18F]F‐Py‐TFP:

[18F]fluoronicotinic acid‐2,3,5,6‐tetrafluorophenyl ester

[18F]AlF:

Aluminum [18F]fluoride

RCP:

Radiochemical purity

EOS:

End of synthesis

t-BuOH:

Tert-butanol

SPE:

Solid phase extraction

TEA:

Triethylamine

TFA:

Trifluoroacetic acid

GC:

Gas chromatography

FID:

Flame ionization detector

TBA-HCO3 :

Tetrabutylammonium carbonate

TLC:

Thin layer chromatography

References

  • Basuli F, Zhang X, Jagoda EM, Choyke PL, Swenson RE. Facile room temperature synthesis of fluorine-18 labeled fluoronicotinic acid-2,3,5,6-tetrafluorophenyl ester without azeotropic drying of fluorine-18. Nucl Med Biol. 2016;43:770–2.

    Article  CAS  Google Scholar 

  • Bouchelouche K, Choyke PL, Capala J. Prostate specific membrane antigen—a target for imaging and therapy with radionuclides. Discov Med. 2010;9:55–61.

    PubMed  PubMed Central  Google Scholar 

  • Cardinale J, Martin R, Remde Y, Schäfer M, Hienzsch A, Hübner S, Zerges A-M, Marx H, Hesse R, Weber K, Smits R, Hoepping A, Müller M, Neels CO, Kopka K. Procedures for the GMP-compliant production and quality control of [18F]PSMA-1007: a next generation radiofluorinated tracer for the detection of prostate cancer. Pharmaceuticals. 2017;10:77.

    Article  Google Scholar 

  • Certificate of analysis for [18F]DCFPyL, personal email communication Windhorst, A.D, VUmc Amsterdam, Netherlands. 25 Dec 2019.

  • Chen Y, Pullambhatla M, Foss CA, Byun Y, Nimmagadda S, Senthamizhchelvan S, Sgouros G, Mease RC, Pomper MG. 2-(3-{1-Carboxy-5-[(6-[18F]Fluoro-Pyridine-3-Carbonyl)-Amino]-Pentyl}-Ureido)-pentanedioic acid, [18F]DCFPyL, a PSMA-based PET imaging agent for prostate cancer. Clin Cancer Res. 2011;17:7645–53.

    Article  CAS  Google Scholar 

  • Dornan MH, Simard JM, Leblond A, Juneau D, Delouya G, Saad F, Menard C, DaSilvia JN. Simplified and robust one-step radiosynthesis of [18F]DCFPyL via direct radiofluorination and cartridge-based purification. J Label Compd Radiopharm. 2018;61:757–63.

    Article  CAS  Google Scholar 

  • Giesel FL, Hadaschik B, Cardinale J, Radtke J, Vinsensia M, Lehnert W, Kesch C, Tolstov Y, Singer S, Grabe N, Duensing S, Schäfer M, Neels OC, Mier W, Haberkorn U, Kopka K, Kratochwil C. F-18 labelled PSMA-1007: biodistribution, radiation dosimetry and histopathological validation of tumor lesions in prostate cancer patients. Eur J Nucl Med Mol Imaging. 2017;44:678–88.

    Article  CAS  Google Scholar 

  • Holland JP, Cumming P, Vasdev N. PET of signal transduction pathways in cancer. J Nuclear Med. 2012;53:1333–6.

    Article  CAS  Google Scholar 

  • Lesniak WG, Mease RC, Chatterjee S, Kumar D, Lisok A, Wharram B, et al. Development of [18F]FPy-WL12 as a PD-L1 specific PET imaging peptide. Mol Imaging. 2019. https://doi.org/10.1177/1536012119852189.

    Article  PubMed  PubMed Central  Google Scholar 

  • Li X-G, Autio A, Ahtinen H, Helariutta K, Liljenback H, Jalkanen S, Roivainen A, Airaksinen AJ. Translating the concept of peptidelabeling with 5-deoxy-5-[18F]fluororibose into preclinical practice: 18F-labeling of Siglec-9 peptide for PET imaging of inflammation. Chem Commun. 2013;49:3682–4.

    Article  CAS  Google Scholar 

  • Lütje S, Franssen GM, Herrmann K, Boerman OC, Rijpkema M, Gotthardt M, Heskamp S. In vitro and in vivo characterization of an 18F-AlF-labeled PSMA ligand for imaging of PSMA-expressing xenografts. J Nucl Med. 2019;60:1017–22.

    Article  Google Scholar 

  • Olberg DE, Arukwe JM, Grace D, Hjelstuen OK, Solbakken M, Kindberg GM, Cuthbertson A. One step radiosynthesis of 6-[18F]Fluoronicotinic acid 2,3,5,6-Tetrafluorophenyl Ester ([18F]F-Py-TFP): a new prosthetic group for efficient labeling of biomolecules with fluorine-18. J Med Chem. 2010;53:1732–40.

    Article  CAS  Google Scholar 

  • Olberg, D.E., Svadeberg, A., 2017. Patent. WO/2017/072200.

  • Pillarsetty, N., Kalidindi, T., Carlin, S., Easwaramoorthy, B., Abbasi, A., Larson, S., Osborne, J., Weber, W., 2016. Effect of specific activity on the uptake of [68Ga]-DKFZ-PSMA11 in tumor and other organs. 57, 528.

  • Ravert HT, Holt DP, Chen Y, Mease RC, Fan H, Pomper MG, Dannals RF. An improved synthesis of the radiolabeled prostate-specific membrane antigen inhibitor, [18F]DCFPyL. J Label Comp Radiopharm. 2016;59:439–50.

    Article  CAS  Google Scholar 

  • Roy J, Kuo F, Basuli F, Williams MR, Wong K, Green MV, Seidel J, Adler SS, Xu B, Choyke PL, Jagoda EM. The distribution volume of 18F-albumin as a potential biomarker of antiangiogenic treatment efficacy. Cancer Biother Radiopharm. 2019;34:238–44.

    Article  CAS  Google Scholar 

  • Zhang K, Bangma CH, Roobol MJ. Prostate cancer screening in Europe and Asia. Asian J Urol. 2017;4:86–95.

    Article  Google Scholar 

Download references

Acknowledgements

The authors wish to extend their gratitude to Anne-Toril Aalvik, Einar Mantor Iversen and Jan Rokne at the Norwegian Medical Cyclotron Centre for isotope production and general technical support.

Funding

This project has been funded in part with federal funds from the South-Eastern Norway Regional Health Authority of Health innovation grant.

Author information

Authors and Affiliations

Authors

Contributions

DEO: Conception and design of the work and aided in the interpretation and analysis of the data. RH: performed the experiments and was a major contributor in writing the manuscript. DEO, RH, TBG and PC were significant contributors to writing and editing of the manuscript and have read and approved the final manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Dag Erlend Olberg.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

DEO is an employee of Norwegian Medical Cyclotron centre Ltd. The fluoride-18 SPE radiolabeling of 18F-Py-TFP labeling agent is patented. The other authors declare no conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hoareau, R., Bach-Gansmo, T., Cumming, P. et al. A new automated and putatively versatile synthesis of the PSMA-ligand derivative [18F]DCFPyL using the FASTlabTM synthesizer. EJNMMI radiopharm. chem. 7, 10 (2022). https://doi.org/10.1186/s41181-022-00157-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41181-022-00157-0

Keywords