Skip to main content

Table 1 Summary of the validation of the quality control methods fo [68Ga]Ga-NeoB

From: GMP-compliant production of [68Ga]Ga-NeoB for positron emission tomography imaging of patients with gastrointestinal stromal tumor

Product identification – HPLC

Selectivity

  Specification

P*

result

9.85 min

  Specification

S*: t(P) - t(S) = 0.05 ± 0.035 min

result

9.80 min

  Specification

V*: t(P) - t(V) > 0.25 min

result

9.48 min

  Specification

G*: t(P) - t(G) > 0.25 min

result

0.68 min

Precision

  Specification

P: mean + 3 s-value

result (n = 6)

9.73 ± 0.20 min

Radiochemical puritiy – HPLC

Resolution R between G and P

  Specification

R > 1.5

result

R = 56.8

Robustness

  Specification

P = 9.73 ± 0.20 min

result

9.77 min

Limit of determination (S/N for 5 MBq/mL)

  Specification

S/N > 10

result

37.9

Recovery

  Specification

W = 80–120%

result

87.3–92.7%

Comparison between RCP HPLC and TLC

 Variance

0.7%

result

HPLC 98.5% TLC 99.2%

Product identification TLC

Precision

  Specification

P Rf = 0.6–0.9

result

Rf = 0.72

  Specification

G Rf = 0–0.1

result

Rf = 0.04

  Specification

P: mean + 3 s-value

result (n = 6)

Rf = 0.72 ± 0.08

Robustness

  Specification

P = 0.72 ± 0.08 min

result

0.69 min

Endotoxine test

 Specification

‘pass’ < 2.5 EU/mL at 1:50

result (n = 6)

all ‘pass‘< 2.5 EU/mL

  1. *P product [68Ga]Ga-NeoB; S non-radioactive standard [69Ga]Ga-NeoB; V precursor NeoB; G generator eluate pure 68Ga3+