Fully-automated production of [68Ga]Ga-PentixaFor on the module Modular Lab-PharmTracer

Abstract

Background

PentixaFor is a promising radiopharmaceutical for positron emission tomography in the detection of different tumor entities and other diseases. Until now, the synthesis of [68Ga]Ga-PentixaFor was reported for the automated synthesis module from Scintomics® only. Our aim was to evaluate the automated synthesis of this radiopharmaceutical on a different module in order to make it available for a broader community.

Results

The synthesis of [68Ga]Ga-PentixaFor with different amounts of PentixaFor (50 μg, 30 μg and 20 μg) on the Modular Lab PharmTracer (MLPT) from Eckert & Ziegler with the already established synthesis template for [68Ga]Ga-DOTATOC yielded best results with 50 μg PentixaFor for clinical multi-dose application. All different quality control parameters tested (e.g. sterility, stability and radiochemical purity) were in accordance with the European Pharmacopoeia.

Conclusions

[68Ga]Ga-PentixaFor was successfully synthesized fully-automated on the synthesis module Modular Lab PharmTracer and can be used for multi-dose application in clinical settings.

Background

In 2011, Demmer et al. first introduced chemokine receptor-4 (CXCR4)-binding peptidic probes for molecular imaging (Demmer et al., 2011). PentixaFor (Synonym for CPCR4.2, Fig. 1) is to date the most promising peptide with a high affinity to chemokine receptor-4. PentixaFor can be used as PET imaging agent when coupled to the positron emitting radionuclide gallium-68. Numerous clinical studies have already been performed in vivo and the interest is steadily increasing in new applications for different tumor entities and other diseases, e.g. inflammatory conditions (Bouter et al., 2018). Examples of clinical applications in oncology are the detection of neuroendocrine tumors, multiple myeloma (Pan et al., 2018), glioma, leukemia (Mayerhoefer et al., 2018) or lymphoma (Luo et al., 2019; Haug et al., 2019). Since the precursor PentixaFor was made available until summer 2018 only for use by the synthesis module from Scintomics®, only hospitals that operated a Scintomics® module were able to prepare [68Ga]Ga-PentixaFor. Here, we would like to introduce the synthesis conditions of [68Ga]Ga-PentixaFor on a different synthesis module, namely Modular Lab PharmTracer (MLPT) from Eckert & Ziegler Eurotope GmbH – in order to introduce [68Ga]Ga-PentixaFor to a broader community. For that purpose, different concentrations of PentixaFor were evaluated based on quality control parameters. Concerning the synthesis of 68Ga-tracers in general, different modules are available on the market which have been reviewed by Boschi et al. In addition, generator post-processing steps like fractionation, anionic-exchange and cation-exchange are described (Boschi et al., 2013). The now proposed fully automated pre-purification of gallium-68 on a strong cation exchange (SCX) cartridge has the advantage that gallium-68 becomes concentrated and germanium-68 as well as non-radioactive impurities (e.g. Zn2+) are trapped on the SCX cartridge.

Fig. 1
figure 1

Structure of PentixaFor (molecular weight: 1221.4 g/mol)

Results

Labeling results of different amounts of PentixaFor

The fully automated production of [68Ga]Ga-PentixaFor was conducted on the commercial labeling synthesis module MLPT. Three different precursor amounts of PentixaFor were tested for radiolabeling with 1080–2500 MBq of gallium-68 (three generators from type GalliaPharm®) - 50 μg, 30 μg and 20 μg. As shown in Table 1, the highest amount of PentixaFor (50 μg) resulted in 94.7% radiochemical yield (RCY, n.d.c.) in the product vial, 2.6% in the waste vial and 1.7% on the C18 cartridge. By lowering the amount of PentixaFor to 30 μg, the RCY of the product decreased to 80.5% (n.d.c.) and the amount on the C18 cartridge increased to 13.3% (n.d.c.). By lowering the amount of starting material to 20 μg, the efficiency became worse. Only 20.0% (n.d.c.) of radiochemical yield was found in the product vial and 69.4% on the C18 cartridge. The distribution of the radioactivity in the cassette for the productions is presented in Fig. 2.

Table 1 Activity measured on different parts of the cassette immediately after synthesis (n ≥ 3)
Fig. 2
figure 2

Overview activity distribution [% of total] of different experiments n ≥ 3 (grey: 50 μg; blue: 30 μg; light blue: 20 μg)

The radioactivity found in the reactor and SCX cartridge did not change significantly by changing the amount of starting material.

In the waste fraction, with decreasing peptide amounts from 50 to 20 μg, the radioactivity increased from 2.6% to 9.3%. Interestingly, the activity measured in the waste vial consists of approximately 25% product and 75% free gallium-68 as seen in the radio-HPLC chromatogram as well as in the radio-iTLC (Supporting Information, Fig. 1).

Quality control of [68Ga]Ga-PentixaFor

The radiochemical purity was evaluated with a standard protocol by radio-HPLC and radio-iTLC.

With radio-HPLC, free gallium-68 would be detected at tR = 2.5 min, whereas gallium-68 bound to PentixaFor was detected at tR = 11 min. Radioactive impurities could not be detected with this method (Fig. 3a).

Fig. 3
figure 3

a Radio-HPLC Chromatogram of [68Ga]Ga-PentixaFor, 50 μg. b UV-HPLC Chromatogram of [natGa]Ga-PentixaFor

With radio-iTLC, no 68Ga-colloide could be detected at Rf = 0.2 but the product at Rf = 0.8 (Fig. 4).

Fig. 4
figure 4

iTLC- Chromatogram of [68Ga]Ga-PentixaFor, 50 μg

The cold standard [natGa]Ga-PentixaFor shows a similar retention time of tR = 11 min as seen in Fig. 3b.

In addition, the product solution was tested for endotoxins. For this purpose, the solution was diluted with endotoxin-free-water in a ratio 1:10. For all samples, an endotoxin concentration below 0.5 IE/mL was detected. This is in accordance with the European Pharmacopoeia (9.0/0125).

With regards to sterility, all products were sterile.

The stability of [68Ga]Ga-PentixaFor in aqueous solution at room temperature was tested up to 4 h via radio-HPLC. As seen in Fig. 5, [68Ga]Ga-PentixaFor is stable at those conditions. No radioactive by-products or free gallium-68 could be detected during this time period.

Fig. 5
figure 5

Stability of 50 μg [68Ga]Ga-Pentixafor

In Table 2, product specifications of the most promising approach with 50 μg PentixaFor are summarized. Acceptance criteria are based on the European Pharmacopoeia (9.0/0125).

Table 2 Summary of the product specifications for 50 μg PentixaFor, n ≥ 3

Discussion

For establishing the automated synthesis of [68Ga]Ga-PentixaFor on the Modular Lab PharmTracer from Eckert & Ziegler Eurotope GmbH for the first time, three different amounts of PentixaFor were evaluated for the production of this tracer with the same standard synthesis template as used for [68Ga]Ga-DOTATOC on the synthesis module MLPT. The results summarized in Table 1 and Fig. 2 demonstrate that the amount of 50 μg of PentixaFor yields the best radioactivity distribution on the cassette. The low standard deviations for 50 μg confirm the high reproducibility of the production.

For 30 μg and 20 μg, an increasing amount of radioactivity on the C18 cartridge and a decreasing yield of activity in the product vial was observed. Theoretically, two scenarios are to our knowledge possible to explain the increased radiochemical yield on the C18 cartridge. Either the product is trapped on the C18 cartridge or uncomplexed 68Ga-colloide. Product could be eluted from the C18 cartridge with pure ethanol. After the automated synthesis is completed, we tried to elute the activity. This resulted in no change of activity measured on the C18 cartridge. As a consequence, no product is stuck on the C18 cartridge. The most likely explanation is that 20 μg and 30 μg are too low to complex 68Ga3+ ions completely resulting in a trapping of unbound gallium-68 in the form of 68Ga3+-colloide on the C18 cartridge.

Based on the analyzation of the waste fraction by radio-HPLC and radio-iTLC, 25% of the activity found in the waste fraction was product. With respect to the total radioactivity used in 50 μg experiments, this means a negligibly product activity loss of under 1%.

The minimum amount needed to fully complex PentixaFor under the described conditions was 50 μg. In former studies in which [68Ga]Ga-PentixaFor was synthesized on a synthesis module from Scintomics®, 40 μg and 20 μg of PentixaFor were used in a final product volume of 15 mL and 14 mL, respectively. The amount of PentixaFor that is safe to be used in vivo is therefore < 20 μg per patient based on those results (Lapa et al., 2016; Wester et al., 2015). In the frame of the here described synthesis protocol, it is important to make sure via standard operating procedures (SOPs), that the maximal injected amount of tracer is below 20 μg/patient. Thus, the maximum injected volume needs to be below 3.6 mL.

With regard to the quality control parameters, we achieved an endotoxine-free, sterile and stable solution of [68Ga]Ga-PentixaFor with a radiochemical purity > 95% over a period of 4 h. Moreover, all tested quality control parameters were in accordance with the European Pharmacopoeia (see Table 2). Calculating the ethanol content based on the parameters chosen in the synthesis template of the MPLT software, results in an ethanol content that cannot exceed 6.2%. Nevertheless, we recommend to either measure the ethanol content of every batch by gas chromatography before releasing it for clinical application, or validate the ethanol content by an external laboratory if a gas chromatography is not available onsite.

Conclusions

[68Ga]Ga-PentixaFor was successfully synthesized fully-automated on the synthesis module MLPT for the first time. All the tested quality parameters for the radiochemical purity, pH, endotoxins and sterility are in accordance with the European Pharmacopoeia. In addition, the product solution is stable for at least 4 h after production, as shown by radio-HPLC. Thus, [68Ga]Ga-PentixaFor can be easily and reliably produced on the module MLPT for clinical application.

Methods

Materials

PentixaFor was obtained from PentixaPharm. An aqueous stock solution of 1 mg/mL was prepared and kept at − 15 °C. All chemicals were of pure chemical grade and solvents for high-pressure-liquid-chromatography (HPLC) were obtained as HPLC grade. TraceSelect water (Sigma-Aldrich) was used in all experiments. The pharmaceutical grade 68Ge/68Ga generator (GalliaPharm®, Eckert & Ziegler Radiopharma GmbH, Germany), Modular Lab PharmTracer (Eckert & Ziegler Eurotope GmbH, Germany) and reagent set EZ-102 (Eckert & Ziegler Eurotope GmbH, Germany) were used. The amount of detected metal impurities/ 68Ge breakthrough as provided by the manufacturer was less than the defined limit in the European Pharmacopeia monograph. Activity counting was determined using a borehole counter (Nuklear-Medizintechnik Dresden GmbH, Germany). HPLC was performed using the HPLC system Knauer Azura (UVD: 2.1 L; P6.1L) coupled with UV and radiometric (Raytest Socket 2″8103 0370) detectors. The TLC scanner used was MiniGita from Raytest. The test for endotoxins was performed using Nexgen PTS (Charles River).

Preparation for labeling of PentixaFor with gallium-68

First, the commercial fully automated synthesis platform MLPT was equipped with a disposable single-use cassette (C4-GA-PEP). The synthesis template and buffer preparations were identical to the ones used for [68Ga]Ga-DOTATOC. All synthesis reagents were contained in the reagent set except the peptide. From the reagent set, 50 mL of 0.9% NaCl and 10 mL of EtOH/H2O (50/50) solution were connected to the cassette on the designated spikes. After connection of the waste vial the preconditioning of the C18 cartridge was performed automatically without user interaction. During the conditioning step PentixaFor was prepared in 20 μg, 30 μg and 50 μg aliquots from the stock solution (1 mg / mL). Meanwhile, the preparation of the acetate buffer solution was performed according to the user manual using EZ 102. The reaction mixture contains 2.2 mL of the final buffer solution, 0.4 mL of EtOH/H2O (50/50) and an aliquot of PentixaFor, which was loaded into the reaction vial. The required volume of eluent was added to the eluent vial prior to synthesis.

Labeling of PentixaFor with 68GaCl3

The synthesis was performed fully-automated without any user interaction. 68Ga3+ obtained from a 1.850 MBq 68Ge/68Ga generator (GalliaPharm®) with TiO2 matrix, was eluted with 0.1 N HCl. The generator eluate was pre-concentrated on a strong cation exchange (SCX) cartridge. [68GaCl4] was recovered from the SCX cartridge by the eluent (5 M NaCl/HCl (0.1 M)) (Schultz et al., 2013). The reaction vial containing the reaction mixture was preheated to 50 °C, and after the 68Ga-activity was transferred to the reaction vial, the temperature was increased to 95 °C. After 5 min at 95 °C, the reaction mixture was cooled down by adding 3 mL of 0.9% NaCl. The crude reaction solution was subsequently transferred to the C18 cartridge for purification. The final product was eluted from the C18 cartridge using a EtOH/H2O (50/50) solution and passed through a 0.22 μm sterile filter into the sterile product vial. The resulting product was diluted to 9 mL with 0.9% NaCl. A sample for quality control was taken.

Quality control

After synthesis, the product vial is removed from the MLPT module and further evaluated for quality control determining the following parameters: total product activity, 68Ga3+-identity via half-life time, chemical purity (pH, sterility and endotoxins) and radiochemical purity (high-pressure-liquid chromatography). The stability of the product at room temperature was monitored by radio-HPLC for 4 h. Radio-HPLC was performed with a standardized method. A: water + 0.1% TFA; B: acetonitrile + 0.1% TFA, gradient: 0–25 min A: 0–100%; 25–28 min A: 100–0%. The limit of detection of the radio-HPLC is 10 kBq/ 20 μL injected volume and the recovery is approximately 80%. For iTLC, ammonium acetate/methanol (1/1) was used as mobile phase and iTLC-SG strips as stationary phase. The sterility tests were performed as described in the European Pharmacopoeia at the Institut für Hygiene, Charité.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Abbreviations

68Ga3+ :

68Gallium3+ ion

68Ge4+ :

68Germanium4+ ion

CXCR4:

Chemokine receptor-4

HPLC:

High-pressure-liquid-chromatography

MLPT:

Modular Lab PharmTracer

PET:

Positron Emission Tomography

RCY:

Radiochemical yield

TFA:

Trifluoro acetic acid

References

  • Boschi S, Lodi F, Malizia C, Cicoria G, Marengo M. Automation synthesis modules review. Appl Radiat Isot. 2013;76:38–45.

    Article  CAS  Google Scholar 

  • Bouter C, Meller B, Sahlmann CO, Staab W, Wester HJ, Kropf S, Meller J. (68) Ga-Pentixafor PET/CT imaging of chemokine receptor CXCR4 in chronic infection of the bone: first insights. J Nucl Med. 2018;59(2):320–6.

    Article  CAS  Google Scholar 

  • Demmer O, Dijkgraaf I, Schumacher U, Marinelli L, Cosconati S, Gourni E, Wester HJ, Kessler H. Design, synthesis, and functionalization of dimeric peptides targeting chemokine receptor CXCR4. J Med Chem. 2011;54(21):7648–62.

    Article  CAS  Google Scholar 

  • Haug AR, Leisser A, Wadsak W, Mitterhauser M, Pfaff S, Kropf S, Wester HJ, Hacker M, Hartenbach M, Kiesewetter-Wiederkehr B, Raderer M, Mayerhoefer ME. Prospective non-invasive evaluation of CXCR4 expression for the diagnosis of MALT lymphoma using [(68) Ga]Ga-Pentixafor-PET/MRI. Theranostics. 2019;9(12):3653–8.

    Article  CAS  Google Scholar 

  • Lapa C, Luckerath K, Rudelius M, Schmid JS, Schoene A, Schirbel A, Samnick S, Pelzer T, Buck AK, Kropf S, Wester HJ, Herrmann K. [68Ga]Pentixafor-PET/CT for imaging of chemokine receptor 4 expression in small cell lung cancer--initial experience. Oncotarget. 2016;7(8):9288–95.

    Article  Google Scholar 

  • Luo Y, Cao X, Pan Q, Li J, Feng J, Li F. (68) Ga-pentixafor PET/CT for imaging of chemokine receptor-4 expression in Waldenstrom macroglobulinemia/lymphoplasmacytic lymphoma: comparison to (18) F-FDG PET/CT. J Nucl Med. 2019;60(12):1724–9.

    Article  Google Scholar 

  • Mayerhoefer ME, Jaeger U, Staber P, Raderer M, Wadsak W, Pfaff S, Kornauth C, Senn D, Weber M, Wester HJ, Skrabs C, Haug A. [68Ga]Ga-Pentixafor PET/MRI for CXCR4 imaging of chronic lymphocytic leukemia: preliminary results. Investig Radiol. 2018;53(7):403–8.

    Article  CAS  Google Scholar 

  • Pan Q, Luo Y, Cao X, Ma Y, Li F. Multiple myeloma presenting as a Superscan on 68Ga-Pentixafor PET/CT. Clin Nucl Med. 2018;43(6):462–3.

    Article  Google Scholar 

  • Schultz MK, Mueller D, Baum RP, Leonard Watkins G, Breeman WA. A new automated NaCl based robust method for routine production of gallium-68 labeled peptides. Appl Radiat Isot. 2013;76:46–54.

    Article  CAS  Google Scholar 

  • Wester HJ, Keller U, Schottelius M, Beer A, Philipp-Abbrederis K, Hoffmann F, Simecek J, Gerngross C, Lassmann M, Herrmann K, Pellegata N, Rudelius M, Kessler H, Schwaiger M. Disclosing the CXCR4 expression in lymphoproliferative diseases by targeted molecular imaging. Theranostics. 2015;5(6):618–30.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

We thank Eckert & Ziegler Eurotope GmbH for technical support. We acknowledge support from the German Research Foundation (DFG) and the Open Access Publication Funds of Charité – Universitätsmedizin Berlin.

Funding

This study was performed within the validation process of [68Ga]Ga-PentixaFor for clinical application and was thus not funded.

Author information

Authors and Affiliations

Authors

Contributions

SS designed, performed and analyzed the experiments. OS designed the evaluation of quality control parameters and repeated the experiments in order to proof its repeatability. WB evaluated the results concerning clinical applications. All authors contributed to, read and approved the final manuscript.

Corresponding author

Correspondence to Sarah Spreckelmeyer.

Ethics declarations

Ethics approval and consent to participate

Not applicable

Consent for publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary information

Additional file 1. Supporting Information, Fig. 1:

iTLC chromatogram of the waste fraction

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Spreckelmeyer, S., Schulze, O. & Brenner, W. Fully-automated production of [68Ga]Ga-PentixaFor on the module Modular Lab-PharmTracer. EJNMMI radiopharm. chem. 5, 8 (2020). https://doi.org/10.1186/s41181-020-0091-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41181-020-0091-2

Keywords