Skip to main content

Table 1 Parameters required for the validation of the HPLC method

From: New sensitive method for HEPES quantification in 68Ga-radiopharmaceuticals

 

Requirements

Results

Retention time

< 4 min

2.8 min

Intra-day

Repeatability

RSD < 5.00%

2.30%

Noise

> 10

43

Inter-day

Reproducability

F-test> 0.05

F-Test > 0.65

Linearity UV-detector

R2 > 0.98

R2 = 0.99

LOD

3x noise

2.74 ± 0.24

LOQ

10x noise

9.40 ± 1.25

Carry over

≤0.10%

0.08%